Installation

Unpacking

Open the shipping cartons. Carefully remove the instrument
and accessories. Lift and carry with two people, holding
securely underneath with both hands. Use proper lifting
technique (lift with the legs, not the back) to avoid personal
injury. Compare the contents with the packing list. If there is a
discrepancy, call Thermo Scientific technical service.
List Of Contents
S No.
Description
1
Thermo Scientific SpeedVac™ SRF110
Concentrator
2
Line Cord
3
RF132 - Hexagonal Micro Centrifuge
Rotor
Inspection
Inspect the unit and accessories for damage that may have
occurred during shipment. Should there be any damage,
report it to the carrier and contact Thermo Scientific
immediately. Make sure the carrier inspects the damage and
leaves an inspection report. Register any claims for shipping
damage against the carrier or his agent. Save the shipping
carton in the event a return is necessary. Call Thermo
Scientific technical service for further assistance.
Site Preparation
The SRF110 requires a stable surface that is clean, dry, level,
and within 4 feet (1.2 meters) of a compatible electrical outlet.
Place the unit in a convenient location with access to a
vacuum source and a vapor trap.
The SRF110 units configured for 115 VAC, 60 Hz, should be
plugged into a circuit rated for at least 20 Amps. The SRF110
units configured for 230 VAC, 50 Hz, should be plugged into
a circuit rated for at least 8 Amps.
SpeedVac™ SRF110 Refrigerated Vacuum Concentrator
CAUTION: For better air circulation, be sure to
leave at least 6-inch clearance on all sides of the
unit. Overhead clearance, equal to 18 inch
(46 cm), is required for raising the cover. Verify
that the unit is on a leveled and stable platform. If
necessary, move the unit to a more suitable
location.
WARNING: Before connecting the unit to an
electrical outlet, make certain that voltage,
frequency, and amperage match the
requirements indicated on the product label,
name plate of the instrument. Use sockets with a
protective earth conductor and correct mains
Qty
cable.
01
Note: Do not use any detachable power cord that is not
adequately rated for the unit.
01
01

Installation of SRF110

Figure 1. SRF110 Installation
CAUTION: Cold trap should be in line between
the vacuum source and the SRF Unit.
Place the SRF110 unit & other components depending upon
on ordered kit and place as shown in Figure 1. Leave at least
6-inch clearance on all sides of the units. This placement
helps to optimize for better refrigeration performance.
Installation |
5