Appendix

Appendix A: Installed configuration Diagrams
Configuration #P1 – Setup with Oil Pump and Cold Trap (High Vacuum System)
Figure 4. Setup with Oil Pump and Cold Trap
Configuration #P2 – Setup with Semi Integrated Vacuum System
Figure 5. Setup with Semi Integrated Vacuum System
1. The relative position of the SRF 110, Cold Trap and
Vacuum Pump should be as shown. There must be at
least 6 inches between the Refrigerated SRF 110 and the
Cold Trap. More space is better. There must be at least 3
inches between the back of the refrigerated SRF 110 and
the adjacent wall surface. This is important to optimize
the air flows of the refrigeration systems. Restriction of
the airflow through the cabinet during operation could
adversely affect performance.
SpeedVac™ SRF110 Refrigerated Vacuum Concentrator
2. After positioning the components, it is necessary to join
the system together using the Tubing Kit provided.
3. Attach one tubing to the tube that extends out the back
of the Concentrator. Attach the other end of this tubing to
one of the barb fittings on the Cold Trap Cover. Secure
the tubing with the clamps supplied.
4. Attach another tubing to the remaining barb fitting on the
Cold Trap Cover and clamp securely. If the accessory
Secondary Trap is not used, attach the other end of the
Appendix |
15