Thermo Scientific SpeedVac SRF110 Installation And Operation Manual page 18

Refrigerated vacuum concentrator
Table of Contents

Advertisement

Problem
Evaporation rate is reduced
Ice formed on Cold Trap lid
Frequent oil change needed
in pump
Unit starts and shuts off
No vacuum/poor vacuum
14 | Troubleshooting
Cause
Heater inoperable
Pump is not functioning
properly
Cold Trap or Concentrator
gasket is not sealing properly
Cold trap lid is not seated
Defrost and wipe dry. the
sealing surface
Secondary Trap Canister not
properly
Secondary Trap Insert is spent
New Secondary Trap has
moisture in it
Secondary Trap Insert is spent
Latch optical sensor improperly
calibrated
Vacuum too strong for chemical
Vacuum pump failure
Obstruction in hose
Lack of adequate vacuum
Pump not on
Control valve open
Leaks in lines or connectors or
gasket
Foreign material on lid gasket
Cold Trap is not emptied after
each run and dried
Corrective Action
Contact Thermofisher.
Check pump by locating vacuum gauge
closer to pump and close off rest of the
system. Checkup pump oil for cloudiness
or particles or change. If pump is faulty,
seek authorized service or replace pump.
Check gasket for cleanliness. Adjust gasket
as needed.
Hold lid down until vacuum is initiated.
Tighten all connections to and fitted from
Secondary Trap Canister.
Replace with New Insert.
Run Vacuum Pump for 24 hours to remove
moisture.
Change insert often.
Unplug Power Cord. Wait 10 seconds. Plug
in power cord.
Use a Secondary Trap Insert and
Diaphragm Pump.
Check pump.
Remove obstruction or replace hose.
See the following No vacuum/poor vacuum
section.
Turn on pump.
Check Control Valve.
Locate and Repair.
Clean Gasket and Lid.
Empty the traps (glass or stainless steel)
after each run and replace.
SpeedVac™ SRF110 Refrigerated Vacuum Concentrator