Appendix D: Chemical
Resistance
SRF110 Centrifugal Concentrator and Cold Trap are
designed to be chemical resistant to most compounds that
are commonly used in concentration processes. SRF110
have several enhancements to minimize chemical attack on
components. The lid, the chamber and rotors are Epoxy
coated, the rotor shaft is Stainless Steel, bearings are Steel,
the coupling insert is polypropylene and all wetted parts in the
vacuum release valves are made from Stainless Steel. Even
though parts in SRF110 are considerably resistant to
corrosion, it is very important that the SRF110 is immediately
cleaned and neutralized.
The following table gives the list of SRF110 compatible
Solvent and Solvent Combination:
Solvent & Solvent Combination
Water + ACN + MeOH
+ low concentration of
1% formic acid
Acetonitrile
Ammonium hydroxide
PCR buffers
(aqueous)
Isobutyl alcohol
n-propyl alcohol
Methanol
CAUTION: If your choice of solvents and
applications are unique and not listed above,
please contact Thermo Fisher Scientific
Customer Support for advice.
Use of non recommended solvent may damage
the product.
WARNING: Do not use this device in radioactive,
highly reactive or explosive atmosphere.
Do not use this device to process any explosive,
radioactive, highly reactive or explosive
atmosphere creating substances.
CAUTION: Product Maintenance:
Maintenance of product is only permitted to be
performed by trained service staff using suitable
measuring and auxiliary equipment as well as
detailed service instruction.
Untrained person can damage the product.
20 | Appendix
Water + ACN + MeOH + low
concentration of 0.1%
Trifluoroacetic acid
0.1% Acetic acid
n-Butyl alcohol
Isopropyl alcohol
Methanol
0.1% Trifluoroacetic acid
Ethanol
Wear gloves, eye protection, masks and lab
coats while working on the system.