Installing the analyzer modules and option module
5.4 Installing the option module OM 2.1
5.4
Installing the option module OM 2.1
1. Isolate the device from power.
2. Remove the housing cover. To do so, unscrew the two nuts on the rear panel of the
3. If applicable, remove the option module blank plate from the rear panel.
4. Remove the printed circuit board holder
5. Insert the option module 2.1
6. Fasten the option module 2.1
Note
Tightening the screws
Tighten the screws on the option module with a torque of
• 1.3 Nm.
Tighten the screws on the PCB holder with a torque of
• 2.5 Nm.
32
device and push the housing cover to the back.
printed circuit board holder
4
①
Option module OM 2.1
②
Printed circuit board holder
Figure 5-3
Installation of option module OM 2.1
②
by loosening the two screws
①
at the appropriate slot into the device.
①
④
with two screws
②
again.
1
2
3
③
④
Operating Instructions, 01/2013, A5E31874006-01
③
to the rear panel and insert the
Screws
Screws
Rack-mounted device
.