Dispose of sample gas through an exhaust gas line
● Keep the flow resistance in the exhaust line small by means of
● The gas line must be free of fast pressure fluctuations. If this is not guaranteed,
Rack-mounted device
Operating Instructions, 01/2013, A5E31874006-01
– A short line and
– A transition to a larger diameter
– Install a special exhaust line or
– Install a damping vessel (> 1 l) with a downstream restrictor between the device and
the exhaust line (pneumatic lowpass filter).
4.1 Operation with at least one OXYMAT 7
Application planning
25