Install four "AA" type batteries into the compartment.
Make sure that the polarities correspond to the (+) and (–)
marks inside the battery compartment.
Batteries can be easily installed or removed as their (-) ends
are pushed against the spring.
You will hear a beep as the batteries are installed.
NOTE: The beep will not sound if AC adaptor is connected.
Close the battery compartment cover. Do not force the battery cover into position.
Insert two tabs to the main unit first, Fig.1). Then close the cover, pressing [
the cover completely Fig.3).
1)
Battery symbol [
] will flash when battery power is low. Measurement can not be made while the
symbol is displayed. Replace the batteries. All four batteries need to be replaced with new ones; do
not mix new and old batteries or different batteries.
The symbol might appear only during measurement although you can review readings in memory. This is
because more battery power is required for conducting a measurement, e.g. inflating the cuff, than
displaying the readings in memory. Please have the batteries replaced.
The enclosed batteries are for monitoring, and their life may be shorter than that of commercial batteries.
2
ADJUST DATE AND TIME AND ACTIVATE THE CLOCK
M A K I N G A M E A S U R E M E N T
Insert AIR PLUG into AIR CONNECTOR before starting a blood
pressure measurement.
The cuff is applicable to arm with circumference between 22 and 32
cm. Take your arm size before measurement.
Choose your memory bank before measurement to use the
personal mode effectively.
Inaccurate readings may result if cuff is
wrapped over heavy clothing or tight shirt that
may restrict circulation in your upper arm or if
sleeve is rolled up.
1
SIT ON A CHAIR AND APPLY CUFF.
Find a chair and a table so that you can be comfortably seated with your feet flat on the floor and your
back and arm supported.
Put the cuff on your left arm with AIR HOSE
positioned toward your hand.
2 to 3 cm
ELBOW
The cuff should fit snugly, leaving enough space for two fingers to be put between
your arm and the cuff.
If the cuff is wrapped more tightly or loosely
than this, inaccurate blood pressure
readings may result.
Adjust the position of the cuff so that AIR HOSE is over the
inner part of your arm over the brachial artery.
Press the surface of the cuff to make sure that the hook &
loop fastens securely.
2
Let your forearm rest naturally on the table and keep the cuff at the height of
your heart during measurement.
If the cuff is lower (higher) than the heart, the measured reading tends to become higher
(lower).
Take deep breaths and relax.
Do not move or speak while taking your blood pressure.
Do not cross your legs during measurement.
Do not hold your breath.
AA (LR6)
BATTERIES
] section Fig.2). Close
2)
PRESS
two tabs
AIR
CONNECTOR
Attach the cuff with hook and loop fastener with
the lower edge of the cuff approximately 2 to 3 cm
above the inner elbow.
AIR HOSE
AIR HOSE
ARTERY
ARTERY
Touch "SET" KEY and keep your
finger on the key until "20 11"
starts flashing.
Clock is set in the order of year,
month, day, hour and minute.
The flashing number increases with MEMORY KEY [1] and decreases with
KEY [2]. The number will be fast-forwarded if you keep your finger on the key.
Touching "SET" key will fix the number and the next article will flash. Touching
"START/STOP" key will terminate the setting.
As the date and time are set, clock is activated.
3)
click!
Display of the clock while the device is turned off indicates that the clock is activated.
OPERATION WITH AC ADAPTOR
AC CONNECTOR is located at the side of the monitor.
Use the designated AC adaptor only.
It is suggested to take out the batteries when using the adapter. However, the clock is inactivated as the
AC adaptor is unplugged and therefore you may insert the batteries to keep the clock activated. In such
case, please make a few measurement without plugging the adaptor to avoid battery leak.
Do not connect the AC adaptor to any power socket where accessibility is not satisfying.
Protection: This device is double insulated and protected against short circuit and overload by a
primary thermal fuse.
Enclosures and Protective Covers: Equipment enclosed to protect against contact with live parts,
and with parts which can become live (finger, pin, hook test).
3
TURN ON THE MONITOR.
Touch "START/STOP" KEY.
Air is exhausted from the cuff.
Automatic inflation starts.
Touch "START/STOP" KEY again to cancel
measurement.
The monitor will exhaust air from the cuff and turn
off.
Inflation symbol disappears and inflation becomes slow.
PULSE SYMBOL is displayed as pulse is detected.
hook and loop
fastener
Air is released from the cuff and measured values are
displayed as measurement completes.
Refer to page 1 for [
Refer to page 1 for [
Refer to page 4 for "Err" display.
4
SELECT MEMORY BANK TO SAVE THE MEASUREMENT RESULT.
MEMORY SYMBOL is displayed below MEMORY KEY.
Touching MEMORY KEY will switch the bank.
Readings are automatically saved in the bank, either MEMORY [1] or [2], selected when the monitor
is turned off after measurement. Make sure to confirm the bank selected after a measurement.
Refer to page 1 for memory function.
5
TURN OFF THE MONITOR.
Touch "START/STOP" KEY.
The monitor will be turned off. Even if you do not turn off the monitor, it turns off automatically after 3
minutes.
SDG!DFG
SGLKSJPREJ
DKLFJGS
OGJOIER;J
Do not execute repeated measurements for congestion of blood could result in false
measurement. Let your arm rest for at least 5 minutes.
-3-
Keep your finger
on the key.
DEFLATION
SYMBOL
INFLATION
SYMBOL
PULSE SYMBOL
PULSE RATE
MEMORY MARK
SYSTOLIC
].
DIASTOLIC
].
WHO CLASSIFICATION
SYMBOL
PULSE PRESSURE
increase
decrease
confirm
AC CONNECTOR
TIME
MEMORY MARK