Nissei DSK-1011J Quick Start Manual page 2

Advertisement

Power Consumption
: 4W (max.)
Electric Rating
: DC6V/4W, with four LR6 batteries
Adaptor; AC100-240V, 50-60Hz, 0.12A,
Unit; DC6V,
Memory
: 2 banks, each stores 60 readings
Operating Condition
: +10˚C to +40˚C, 30% to 85% RH (noncondensing)
Transportation/Storage Condition : -20˚C to +60˚C, 10% to 95% RH (noncondensing)
Cuff
: Coverage arm circumference; 22 to 32 cm, applied part; BF
Main Unit
: Weight; Approx. 250 g (without batteries)
Size; Approx. 115 x 115 x 65.9 mm (W x D x H)
Protection against harmful ingress of water
: IPX0
P R E C A U T I O N S A N D C O N T R A I N D I C A T I O N S
Read this instruction manual carefully before use and use this product correctly. The instructions below
are to help you use the product correctly and to prevent you from being harmed or injured.
The subjects or use which exceeds the responsibility of the manufacturer such as the use beyond
Prohibition
the design limit or the unjustified use. (Ones with possible death or serious injury are indicated.)
General cautions for use of the product. (The subjects which may result in personal injury or
Caution
property damage in case the product is incorrectly handled are indicated.)
*Property damage means consequential damage for house, household belongings, livestock or pet.
• Do not use the product for infants or persons who cannot express their will. It may cause an
accident or an injury.
• Do not bring the product in the test room when having a MRI scan. It may cause the attraction
by magnetic force or the thermal burn.
• Do not use the product near flammable gas, such as anesthetic gas, or in locations or
atmospheres with high oxygen concentration such as hyperbaric oxygen chambers/therapy
devices or oxygen tents. It may cause malfunction, breakage, deterioration or inducement of
explosion.
• Do not make any diagnoses or treat yourself based on your own measurement results. Make
measurement based on the consultation with your doctor and receive professional treatment and
prescription.
• Do not use the product, if you have an injury on/around your arm. It may cause the
aggravation of the injury.
• Do not use the product during intravenous drip injections or blood transfusions. It may
cause an accident or an injury.
• Do not use the product with exceeding its service life. The service life of the product is either 5
years or 30,000 times.
• Do not use the product in the medical institution or public area where the product is
subject to be used by unspecified persons. It may cause an accident or a trouble.
• Do not use the product near mobile phones, micro ovens or other strong electromagnetic
devices. It may cause malfunctions or damage the product.
• Be cautious about using the product in conjunction with the medical electrical equipment
listed below. It may cause malfunctions.
· Pace makers, implantable defibrillators, or other wearable medical electrical equipment that is
easily affected by electromagnetic force
· Wearable medical electrical equipment such as an electrocardiographic monitor.
• Consult a doctor when persons in pregnancy use the product.
• Do not use the product for purposes other than measuring blood pressure. It may damage
the product or cause an accident.
• Do not use the cuff around anywhere other than your upper arm. It may damage the product
or cause an accident.
• When using the monitor in an environment where systems connecting fluid with blood
T I P S O N T A K I N G Y O U R B L O O D P R E S S U R E
Blood pressure is the force exerted by the heart in pumping the blood through the arteries and the
resistance by the veins in this flow.
Blood pressure varies all the time, influenced by mental and physical factors, and is never
constant.
In general, blood pressure is highest during the working hours and gradually decreases during the afternoon
and evening hours. It is low during sleep and increases at a relatively fast rate after arising from bed.
Causes for Changes in Blood Pressure
· Time and season
· Physical activity
· Eating and drinking (including Alcohol)
· The urge to urinate
· Conversation
· Measurement posture (including the cuff level against the heart)
Let a qualified physician interpret your blood pressure readings.
Depending on your age, weight and general condition, blood pressure can be slightly different. Consult
with your doctor on determining what blood pressure is normal for you.
P R E P A R A T I O N B E F O R E U S E
Install batteries and activate the clock before use. Measurement can be made with clock being
inactivated. However, activate the clock to save measurement date and time with readings for better
review. The clock is inactivated as the batteries are removed. The product can be also operated with
AC adaptor. See page 3.
500mA, with designated AC adaptor
Prohibition
Caution
· Taking medicine for high blood pressure
· Smoking
· Mental nervousness
· Other environment (during medical examination etc.)
Protection against electric shock : Internally powered equipment, Type BF applied part (cuff)
Mode of operation
Classification
Key to symbols
Specifications are subject to change without prior notice.
This product complies with JIST1115 (noninvasive electric blood pressure monitors) and EMC
(electromagnetic compatibility) Standard IEC60601-1-2:2007. (CISPR Classification: Group 1, Class B)
* Test procedure is based on ANSI/AAMI SP10:2002, American National Standard for electronic or
automated Sphygmomanometers, and Amendment.
vessels such as dialysis and intravenous drip injection are used, do not connect the blood
pressure monitor and the system. An accident may be caused if the air enters a blood vessel.
• Consult a doctor before using the product for persons suffering from arrhythmia or peripheral
circulatory disorders caused by diabetes, liver disease, arteriosclerosis or hypertension.
• When using the cuff, use the supplied equipment. Using equipment other than that supplied
may cause measurement error.
• Measure the circumference of your upper arm and make sure that it is within the suggested
range. Using the product for an arm outside the suggested range may cause measurement error.
• Do not place heavy items on the product or tie up the air hose. It may damage the product.
• Do not make an impact on or drop the product. It may cause malfunctions or damage the product.
• Wear the cuff correctly and keep it level with the heart during measurement. Wearing the cuff
incorrectly or at the wrong height may cause measurement error.
• Be cautious that blood pressure is affected by the factors below.
· Time and season
· Eating and drinking (including Alcohol)
· The urge to urinate
· Measurement posture (including the cuff level against the heart)
• Touch the "START/STOP" KEY if the product applies an excessive level of pressure to your
arm. Touching the "START/STOP" KEY releases the air from the cuff immediately, and the
product stops measuring.
• The cuff may leave bruises on your arm. If you feel any pain, touch the "START/STOP" KEY
to stop the operation of the product. Touching the "START/STOP" KEY releases the air from
the cuff immediately, and the product stops measuring.
• Do not plug/unplug the AC adaptor with wet hands, or touch it with tweezers, or other
electronically conductive devices.
• Do not touch another person while touching the batteries, battery terminals of the device or
the output plug of the AC adaptor, such as when replacing the batteries or
plugging/unplugging the AC adaptor.
• Do not use new batteries in conjunction with old batteries or different types of batteries at
the same time. It may generate heat and damage the product.
• Pull the plug out from the power outlet if anything unusual is detected during use. Plugging
the AC adaptor in a power outlet located behind furniture may be inconvenient if it needs to be
pulled out in an emergency.
• Do not unplug the AC adaptor by pulling the cable. It may damage the cable and cause a fire
or electric shock.
• Take the batteries out when not using the product for a long period of time. Battery fluid may
leak from the batteries and damage the product.
• Do not allow children to handle the product alone or leave the product in a place children
can reach. It may injure the children.
• Do not disassemble, repair or modify the product. It may cause malfunctions or damage the product.
• Check if the product is dirty or wet before use.
• If the product stops functioning properly, stop using it. Please contact us for repair or
assistance.
• Follow the local regulations of your area when disposing the product or batteries.
Blood pressure measured at home tends to be lower than when measured in a hospital,
clinic or doctor's office.
This is because you are tense at the hospital and relaxed at home. It is important to know your stable
normal blood pressure at home.
Before taking blood pressure, rest for approximately five minutes and take your blood pressure while
relaxed in a quiet environment.
Measure blood pressure using the correct posture and do not move or speak during measurement.
· Bathing
Avoid exercise, eating, drinking alcohol, smoking and other activities that affect your blood pressure
right before a measurement.
Take your blood pressure at the same time every day.
The ambient temperature should be approximately 20˚C when you take your blood pressure.
Reading may vary slightly, depending on posture, whether sitting or lying.
1
INSTALLING BATTERIES
Open the battery compartment cover, pressing the projection
between [
] and [
-2-
: Continuous operation
: Internally powered equipment
: Type BF applied part
: Refer to instruction manual/booklet.
: Class II equipment
· Physical activity
· Taking medicine for high blood pressure
· Smoking
· Mental nervousness
· Conversation
· Other environment (during medical examination etc.)
] at the bottom of the main unit.
· Bathing