Performing a Control Solution Test
For example only:
If you are using the Level 2 control
solution, according to this particular vial
of strips, your meter should show a
number between 80 and 120
Note:
- Element™ Brand control solution is sold separately.
- It is recommended that glucose control solution is stored at room temperature
68~77℉(20~25℃) before testing.
- Check the expiration date before performing a glucose control solution test.
- Do not used if expired. Once opened, solution expires after 3 months.
If your control solution test falls out of range, please follow the next steps before
contacting customer support:
• Check the expiration dates on all the products you are using. Be sure you are
using Element™ Brand control solution.
• Try another control solution test.
If this test falls out of range, try another control solution test with a new unopened
bottle of strips:
• After following the appropriate steps and the control test still falls out of range, do
not perform a glucose test.
Please contact an Infopia USA representative at 1- 888-446-3246.
Caution:
• If your glucose control solution falls out of range, do not perform a glucose test, please
contact your local representative.(1-888-446-3246 or www.infopiausa.com)
• The Glucose Control Solution range is not the recommended range for your blood
glucose level.
• Discard the used control solution and test strips carefully according to the local
regulations.
22 / 23
www.infopiausa.com
Level 1. 35-65
Level 2.
80-120
Level 3.
255-345

Performing Your Test

Note: If during the setup process you chose to use the voice feature, your
Element™ Plus meter will announce the following steps and results in
the designated language setting, English or Spanish.
Step 1
Firmly insert the test strip into the meter
port with the "Blue Color" facing up. The
Infopia logo should be facing towards you.
The meter automatically turns on display-
ing the code, room- temperature, date and
time.
Make sure the code number matches the
code on the strip bottle. The Element™
Plus automatically recognizes the test strip
code number and adjusts the meter
accordingly.
If during the setup process you chose to use
the activity/meal flags (
and release
activity that will correlate with your result,
then press and release the power button.
" The code number is twenty one."
and
button to select the
" Please choose your activities ."