Clicking on any of the file names will open up the data file. These files are CSV files and typi‐
cally open into an Excel spreadsheet. Once open, the spreadsheet can be saved to another
location such as a hard drive. A typical entry would look like the file below.
(HR:MIN:SEC) (Count) (Count)
The TIME column indicates the specific time, using a 24 hr clock, the count data was taken.
The second column indicates the RAW COUNT, while the third column is the ENGINEERING
COUNT. The RAW COUNT divided by the C/W RATIO equals the ENGINEERINING COUNT. The
next two columns are both measurements of the full scale, but shown in different units. The
unit (% Volts) is used in the Static Calibration procedure. The units (% PV) or Process Variable,
is the calibrated full scale measurement. The two units, (% Volts) and (% PV) are mathemati‐
cally related by the CALd multiplier. The RATE column indicates the flow rate in the mass and
time units chosen. The final column, ALARM, indicates whether or not the alarm has been
triggered. (Note: The (% PV) column is recorded as an integer with an implied decimal point.
Divide the (% PV) column by 100 in order to get the correct real number.)
PLESE NOTE THAT THIS LOG HAS BEEN DISABLED IN SOFTWARE VERSION 8.5.0. IF YOU
WOULD LIKE IT TO BE ENABLED, PLEASE CONTACT EASTERN INSTRUMENTS.
416 Landmark Drive
Wilmington, NC 28412
TIME
RAW
ENGR
14:55:30
50
0.27778
14:55:31
183
1.0167
14:55:32
318
1.7667
14:55:33
435
2.4167
14:55:34
560
3.1111
14:55:35
673
3.7389
14:55:36
823
4.5722
14:55:37
935
5.1944
14:55:38
1061
5.8944
14:55:39
1199
6.6611
www.easterninstruments.com
Volts
PV
RATE
(%)
(%)
(Mass/Time) (Over)
18.18
1817
1817
27.46
2745
2745
23.9
2390
2390
25.52
2551
2551
25.13
2512
2512
24.97
2497
2497
25.49
2548
2548
24.98
2497
2497
25.13
2513
2513
25.41
2540
2540
38
ALARM
0
0
0
0
0
0
0
0
0
0