Step 6:  Hang  the  Test  Weight  provided 
with the Electronics Enclosure from 
the  Calibration  Stud  located  under 
the bottom edge of the Pan as seen 
below.    For  Pans  that  have  more 
than  one  Calibration  Stud,  there 
should  be  a  Test  Weight  for  each.  
Some optional pans may not have a 
Stud.    In  these  cases  the  wire  will 
be  long  enough  to  hook  onto  the 
top edge of the pan and be draped 
over  the  front  end  of  the  pan  as 
shown  below.    Be  sure  that  the 
Test Weight hangs freely from the Calibration Stud.  It is also good practice to allow 
the Test Weight to become stable and movement free before proceeding. 
 
Step 7:  Press the CALs Button located on the lower left side of the screen.  Note the % Volts 
 
reading  displayed  in  the  Diagnostic  Window.    Verify  the  reading  is  correct  for  the   
 
selected range.  For example, 25% is the correct value for the Standard Range setting.  
 
Please see the table below for the correct % Volts reading for each range.  
 
 
 
 
Step 8:  Remove  the  Test  Weight  from  the  Calibration  Stud.  The  voltage  should  return  to 
 
0.00% Volts.  
416 Landmark Drive 
Wilmington, NC 28412 
Range
Extended Standard Reduced Minimum
% Volts
12.5%
PAN WITH PAN LINER 
AND CALIBRATION STUD 
www.easterninstruments.com 
25.0%
50.0%
100.0%
25
 
ALL‐WELDED PAN LINER 
WITHOUT A CALIBRATION STUD 