2,200 mL
Stopped
Taper
1:15 PM
Adult
Adult
TPN
TPN
mL
m
CADD
Infusion Volume
2,000
mL
Taper Up
1
Hrs
0
Min
Taper Down
1
Hrs
0
Min
Infusion Duration
12
Hrs
0
Min
'Accept Value', or press 'select' to edit
Alarm Help Screens
Exit
Accept Value
When certain alarms occur, additional information may be displayed on the screen.
The help screens describe what you can do to try to solve the problem that is causing the alarm.
Steps:
1. When an alarm occurs, press SILENCE to quiet the alarm.
2. If help screens are available for the alarm, "Help" appears above the right soft key.
3. Follow the applicable instructions provided on the help screen. To page through all of the
4. "Retry Help" appears when no additional help steps are available. To page through the help
The following alarms have associated help screens. Refer to steps above to navigate the help screens.
Cannot start pump without a latched cassette.
Alarm
Cannot start pump
without a latched
cassette.
Acknowledge
Silence
Upstream occlusion. Clear occlusion between pump and reservoir.
Alarm
Upstream occlusion.
Clear occlusion
between pump and
reservoir.
Stop Pump
Silence
Alarm - Upstream Occlusion
Reposition bag in pouch
Stop Pump
Next
-Solis VIP Ambulatory Infusion System
®
To view the help screens, select HELP.
available help screens, press NEXT repeatedly. Press ACKNOWLEDGE at any time to exit Help.
screens again, press RETRY HELP. The alarm screen reappears as in Step 2 above. Repeat
steps 2 and 3. When the alarm clears, the help screen will no longer be displayed.
Alarm
Cannot start pump
without a latched
cassette.
Acknowledge
Alarm
Upstream occlusion.
Clear occlusion
between pump and
reservoir.
Stop Pump
Alarm - Upstream Occlusion
Call your home care
provider.
Stop Pump
Retry Help
Alarm - Cassette Not Latched
Latch cassette.
Help
Acknowledge
Alarm - Upstream Occlusion
Fully insert spike.
Help
Stop Pump
Alarm - Cassette Not Latched
Call your home care
provider.
Next
Acknowledge
Alarm - Upstream Occlusion
Unkink tubing.
Next
Stop Pump
Retry Help
Next
Need help?
Do you have a question about the CADD-Solis VIP and is the answer not in the manual?
Questions and answers