Entering KV vs. SS Tables

1-54
The KV/SS submenu enables you to enter a table of values of the
kinematic viscosity (KV) based on signal strength (SS). You can then
use this table to calculate KV as shown in the KV/SS option (in the
ADVAN option of the SETUP submenu) discussed on page 1-39. To
use the option, you must enter the KV vs. SS pairs in descending
order of KV and ascending order of SS. For example, if pair 1 is KV
= 10 centistrokes and SS = 62, then pair 2 must be KV ≤ 10
centistrokes and SS ≥ 62. The DF868 accepts 2 to 20 pairs.
KV/SS
1. Enter the
submenu by pressing [→] twice and
initial User Program screen.
[ENT]
2. Press
to accept the current # of KV/SS Pairs or enter a new
value (from 2 to 20) and press
[ENT]
3. Press
to accept the current Sig Strength value or enter a new
value (50.0 to 85.0) and press
values in ascending order.
[ENT]
4. Press
to accept the current Kin Viscosity value or enter a new
value (0.050 to 500.) and press
viscosity values in descending order.
The Sig Strength and Kin Viscosity prompts repeat for each pair. After
entering all the pairs, the meter returns to the Program menu as
shown on page 1-8. To leave the User Program, press
[F2]
[ENT]
.
[ENT]
. Enter the signal strength
[ENT]
. Enter the kinematic
[EXIT]
Programming Site Data
at the
.