Condensed Operating Instructions
Saving Setups
1. To save the current set of test conditions as entered (conditions include, voltage, test
times, measurement range, limit, and display modes):
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Utilities menu.
Press Up or Down Arrow key to select Save Setup.
Press [ENTER] key to activate setup entry field.
Press UP or Down Arrow key to select the desired setup, choices are:
- NEW, if a setup is to be saved in instruments internal memory.
- DEFAULT, if conditions at power up are to be changed and restored.
- FLOPPY, if a new setup is to be stored or existing one changed on the
floppy.
- Setup Name if an existing set of conditions are to be changed in the
instrument internal memory or floppy.
Press [ENTER] key to activate the Save entry field selected above.
If NEW is selected (internal memory or floppy) enter the desired identifying
name, up to 8 characters, under which these set of test conditions will be stored.
Press [ENTER] key to finalize storage of the setup conditions or if DEFAULT
is selected one must answer Y or N to overwrite.
Changing Test Voltage
1. Set the desired test voltage:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Setup menu.
Press Up or Down Arrow key to select Voltage.
Press [ENTER] key to activate the voltage entry field.
Enter desired test voltage (up to 5 digits with decimal) between 1 and 1000V.
Press [ENTER] key to finalize the voltage entry.
Changing Measurement Times
1. To set the desired measurement cycle times, charge, dwell, measure and discharge:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Setup Menu.
Press Up or Down Arrow key to select desired time: charge, dwell, measure or
discharge.
Press [ENTER] key to activate the corresponding time entry field.
Enter the desired time between 0 and 300 seconds(measure, 0 – 999 on later units)
Press [ENTER] key to finalize the time entry.
2. Continue to set the other times as desired by repeating step 1.