QuadTech 1865 Instruction Manual page 18

Megohmmeter/ir tester
Table of Contents

Advertisement

Condensed Operating Instructions
Saving Setups
1. To save the current set of test conditions as entered (conditions include, voltage, test
times, measurement range, limit, and display modes):
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Utilities menu.
Press Up or Down Arrow key to select Save Setup.
Press [ENTER] key to activate setup entry field.
Press UP or Down Arrow key to select the desired setup, choices are:
- NEW, if a setup is to be saved in instruments internal memory.
- DEFAULT, if conditions at power up are to be changed and restored.
- FLOPPY, if a new setup is to be stored or existing one changed on the
floppy.
- Setup Name if an existing set of conditions are to be changed in the
instrument internal memory or floppy.
Press [ENTER] key to activate the Save entry field selected above.
If NEW is selected (internal memory or floppy) enter the desired identifying
name, up to 8 characters, under which these set of test conditions will be stored.
Press [ENTER] key to finalize storage of the setup conditions or if DEFAULT
is selected one must answer Y or N to overwrite.
Changing Test Voltage
1. Set the desired test voltage:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Setup menu.
Press Up or Down Arrow key to select Voltage.
Press [ENTER] key to activate the voltage entry field.
Enter desired test voltage (up to 5 digits with decimal) between 1 and 1000V.
Press [ENTER] key to finalize the voltage entry.
Changing Measurement Times
1. To set the desired measurement cycle times, charge, dwell, measure and discharge:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select Setup Menu.
Press Up or Down Arrow key to select desired time: charge, dwell, measure or
discharge.
Press [ENTER] key to activate the corresponding time entry field.
Enter the desired time between 0 and 300 seconds(measure, 0 – 999 on later units)
Press [ENTER] key to finalize the time entry.
2. Continue to set the other times as desired by repeating step 1.