Default Measurement Conditions - QuadTech 1865 Instruction Manual

Megohmmeter/ir tester
Table of Contents

Advertisement

In the manual mode the test cycle timing is totally at the users discretion where each of
the three phases, charge, measure or discharge is initiated directly by the user. This mode
would typically be used in an engineering environment or component evaluation where
the measurement results can be observed as test cycle and measurement conditions are
altered.
Whenever the 1865 unit is powered up it is ready to begin measuring at default test
conditions. Initially, these conditions will be set to a factory default (1V, 0 times,
Resistance Display, Auto ranging and Automatic Mode), but can be changed by the user
and stored to overwrite factory default.

2.5.2 Default Measurement Conditions

A set of default measurement conditions are initially established at the factory and stored
in instrument memory. Default conditions are those that determine the instruments status
on power up, thus the instrument is always set to a known state before any testing begins.
These conditions can be changed by the user for tailoring to a specific application. Refer
to paragraph 2.6.4.1 under Save Setup on the Utilities menu.
Factory default measurement conditions are:
Under Setup Menu
Voltage - 1V
Charge, Dwell, Measure and Discharge times – 0
Mode - Auto
Range - Auto (Selected on lower level menu)
Limit - None
Stop on Pass - No
# to Average - None
Under I/O Menu
Display Type - Resistance (selected on lower level menu)
Result Format - Engineering Units
RS-232 - Enable
Handler - On
Under Utilities Menu
Lockout - Off
Backlite - On
Operation
Page 33 of 87