2.8 Error Messages

"BAD DRIVE-REQUEST STRUCTURE"
"BAD VOLTAGE SENT TO FUNCTION"
"BAD ZERO CALIBRATION"
"CALIBRATION STANDARD OUT OF
RANGE"
"CORRUPT SETUP DATA"
"CRC ERROR ON FLOPPY"
"DISK WRITE PROTECT ERROR"
"ELAPSED TIME ERROR, SET TO 0"
"FAILURE # SEE DOCUMENTATION"
"FILE ACCESS VIOLATION"
"FILE EXISTS"
"FLOPPY DRIVE NOT READY"
"FLOPPY MEDIA DEFECT"
"FLOPPY READ FAULT"
"FLOPPY SECTOR UNFORMATTED"
Page 70 of 87
Floppy option not present, defective or disk
not inserted
Improper voltage entered during calibration
procedure
Zeroing error, repeat and/or remove
component from input terminals/fixture
(Refer to Note on p.63)
Entered value > than 1% of 1 Megohm
Stored setups lost or corrupt (battery for
non-volatile memory may be defective or
has been disconnected)
Stored setup lost or corrupt (cyclic-
redundancy-check)
Floppy disk is write protected
Elapsed time is incorrect and has been set
to zero (battery for non-volatile memory
may be defective or has been disconnected)
Hardware or software failure, 1 - 99
(# list not included in this manual)
File selected from floppy is Read Only
File already exists under the chosen name
Floppy option not present, defective or disk
not inserted
Disk defective, format or replace
Floppy option not present, defective or disk
not inserted
Floppy disk is not formatted
Operation