Cpr Procedure - Invacare Softform Premier Active 2 User Manual

Hybrid mattress
Hide thumbs Also See for Softform Premier Active 2:
Table of Contents

Advertisement

Pos.
Description
Power failure indicator
D
E
Pressure failure / Low pressure
indicator
Operation
1. The power indicator will flash and control unit will enter static mode initially until pressure reaches 15mmHg. Then control
unit will enter alternate mode and power indicator will remain on. The cycle time is set at 10 min, and the pressure is
set at 60±3mmHg.
The first inflation of new mattress: the indicator function (visible and audible indicators) will activate if the mattress does
not inflate completely within 15 minutes.
Operation failures
During control unit operation, low pressure indicator will activate within 1 minutes if air hoses are disconnected or
air cells are broken.
In the case of abnormal alternate or no alternate, the audible signal will be activated and alternate failure indicator will flash.

5.3 CPR procedure

The Invacare® Softform Active® 2 Rx mattress has been
fully tested to comply with the current CPR standard of 5-
6 cm compression depth. This was achieved at all stages
of inflation/deflation.
1632862-A
Function
Red LED indicator flashes and audible buzzer sounds if the system has
no power connected. This could be in the event of a power failure, for
example caused by power cord unplugged or power off when control
unit is working. When power is restored, control unit will automatically
start working.
Red LED indicator flashes and audible buzzer sounds if the system
recognises a low pressure failure.
Usage
15