Invacare Softform Premier Active 2 User Manual page 23

Hybrid mattress
Hide thumbs Also See for Softform Premier Active 2:
Table of Contents

Advertisement

Cause
Problem / Alarm
Power outage.
Fuse blown.
Disconnection of CPR (connection hose).
Pressure failure / Low
pressure indication
Disconnection of connector tubes to air cells in
air insert.
Kinked connection hose
In case of issues with troubleshooting, please contact Invacare for further assistance (contact details on the back
page of this User Manual).
1632862-A
Solution
Wait until the power source has power.
Change fuse on power inlet connector with spare
fuse or identical replacement only (consult a
trained engineer if you are unsure how to change
a fuse).
Connect hose properly.
Check individual air cells in insert are correctly
connected to connector tubes.
Make sure that there are no kinks or bends in
the hose.
Troubleshooting
23