Invacare® Softform Premier Active 2 with RX pump
Immunity test
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Immunity test
IEC 60601 test level
3 V
Conducted RF
rms
IEC 61000-4-6
150 kHz to 80 MHz
6 V
rms
28
IEC 60601 test level
Compliance level
30 A/m
30 A/m
Compliance level
3 V
rms
150 kHz to 80 MHz
in ISM bands
6 V
rms
Electromagnetic environment – guidance
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the CT515,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = [3.5/V
in ISM bands
] x √P
1
1632862-A