Summary of Contents for Invacare Softform Premier Active 2
Page 1
Invacare® Softform® Premier Active 2 and Active Care en Mattresses User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
This is to certify that your Invacare® Softform®Mattress is warranted symbols are accompanied by a heading that indicates the severity by Invacare Ltd for a period stated in the Table “Technical Data” of of the danger. this user guide. The Warranty of your Invacare® Softform® product is valid from time of shipping.
General 1.5 Intended use If a defect or fault is discovered the Invacare® customer or Local Business Development Manager from whom the appliance was This pressure redistribution mattress and control unit are intended obtained must be notified immediately. The manufacturer will not...
Invacare® Softform® Premier Active 2 and Active Care 2 Safety WARNING! Invacare® products are specifically designed and manufactured for use in conjunction with Invacare® 2.1 Safety information accessories. Accessories designed by other manufacturers have not been tested by Invacare® and WARNING! are not recommended for use with Invacare®...
– Check all parts for shipping damage and test before using. – In case of damage, do not use. – Contact Invacare® / Carrier for further instructions. 2.2 EMC information There is no known risk of adverse effects on the Active 2 / Active Care control unit caused by other electromagnetic devices, present at the time of treatment.
Invacare® Softform® Premier Active 2 and Active Care 3 Components U-shape, non-castellated base layer Alternating air cell insert 3.1 Product description Castellated insert The Softform®Premier Active 2 / Active Care mattress system Micro-processor controlled control unit acts as a static pressure reducing support/mattress for patients at...
Setup 4 Setup WARNING! Risk of entrapment! Patient entrapment with the bed side rails may cause 4.1 Safety information injury or death. A thorough patient assessment should be completed and monitored and the equipment should WARNING! be used as specified and maintained to reduce the risk Electrical shock hazard! of entrapment.
Invacare® Softform® Premier Active 2 and Active Care 4.2 Installing the system WARNING! This mattress is recommended to be installed on medical 1. Place the pump unit at the end of the bed frames with bed sides or assist rails. It is preferred bed using the built in pump hooks.
Usage 5 Usage IMPORTANT! If holes are present in the mattress cover, there is a danger that liquids may ingress and contamination may 5.1 Safety information occur. – Medical equipment including infusion pumps and Remove all packaging before use. monitors should be attached to appropriate bed Place the mattress directly on the frame of the bed.
Invacare® Softform® Premier Active 2 and Active Care 5.2 CPR procedure The Softform®Premier Active 2 / Active Care mattress has been fully tested to comply with the current CPR standard of 4 – 5 cm compression depth. This was achieved at all stages of inflation/deflation.
Usage 5.3 Control unit Software Schematic Switch on pump System Setup English Language System OK Alarm event Pressure 60mmHg Français Rotor Failure 40 50 60 70 80 Deutsch Español Low Pressure Italiano Português Svenska Dansk Norsk 1576248-C...
Invacare® Softform® Premier Active 2 and Active Care 5.4 Control unit Menu Display LCD display screen Up arrow button OK button / Silence alarm Down arrow button Green LED light Task Action Display 1. Power up Push main switch of control unit.
Page 15
Usage Task Action Display 3. How to enter menu Press arrow Up. “System OK” with “1” underneath on LCD display. then Press arrow Down. “System OK” with “2” underneath on LCD display. then Press OK. “Pressure” displayed on LCD display. (3 second time limit between each function).
Invacare® Softform® Premier Active 2 and Active Care 6 Transport 6.1 Safety information IMPORTANT! – Take care when handling mattresses to ensure no damage to the cover. It is recommended that two people life/carry mattresses. – Avoid contact with jewellery, nails, abrasive surfaces etc.
Maintenance Cleaning control unit 7 Maintenance WARNING! 7.1 Inspection (multiple use) Electrical shock hazard! It is recommended that the mattress is checked (foam, air insert and – Ensure the control unit is disconnected from the cover) for strike-through (this may include fluid ingress, stains, rips or mains electricity supply before cleaning.
Page 18
Invacare® Softform® Premier Active 2 and Active Care Drying covers Heavy soilage Where the mattress is badly soiled, we recommend cleaning with a Hang mattress covers from a line or bar and drip dry in a clean dilute cleaning solution at 80 °C in the washing machine.
Page 19
Maintenance IMPORTANT! – Do not use phenols, alcohols, bleaches, or other abrasive materials. Replacing covers Unzip the cover and remove it carefully from the foam core. Place new cover onto the foam core. Then close the zipper. IMPORTANT! – Ensure that the corners of the foam core are positioned correctly into the corners of the cover.
Invacare® Softform® Premier Active 2 and Active Care 8 After Use 8.1 Storage IMPORTANT! – Store mattresses in a dry environment. – Store mattresses within a protective cover. – Store items on clean, dry, off-flooring free from sharp edges to avoid any possible damage.
Mattress CPR hose disconnected. Connect CPR hose connector, lock it in place. (not alternating Has power and fuse is good, control unit does not come Send control unit back to Invacare for repair. properly). Major leak in air cell. Replace leaking air cell.
Page 22
If the fault reoccurs after the system has been through set up, the pump alarm will sound again, to notify the user of the fault. In case of issues with troubleshooting, please contact Invacare for further assistance (contact details on the back page of this User Manual). 1576248-C...
Technical data 10 Technical data 10.1 General Data Product Warranty Firetesting Nominal Nominal Maximum user Grade ref & Weight of density range hardness weight [kg] Colour product [kg/m3] range [N] [kg] SOFTFORM® Cover: 4 years EN 597-1 RX 39/120 Pink 38 –...
10.5 Spare parts Operating Pressure 60 mmHg (8 kPa) For spare parts/components list, please contact your nearest Power 23 VA Invacare dealer in your country specified on back of this user manual. Control unit fuse T1 AL 250 V 10.3 Materials Foam...
Need help?
Do you have a question about the Softform Premier Active 2 and is the answer not in the manual?
Questions and answers