Immunity test
IEC 60601 test level
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
385MHz - 5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
1632862-A
Compliance level
Electromagnetic environment – guidance
d = 1.12 x √P 80 MHz to 800 MHz
10 V/m
80 MHz to 2.7 GHz
d = 2.3 x √P 800 MHz to 2.5 GHz
385MHz - 5785MHz
Where P is the maximum output power rating of the
Test specifications
transmitter in watts (W) according to the transmitter
for ENCLOSURE
manufacturer and d is the recommended separation
PORT IMMUNITY
distance in meters (m).
to RF wireless
Field strengths from fixed RF transmitters, as determined
communication
by an electromagnetic site survey
equipment (Refer
the compliance level in each frequency range
to table 9 of IEC
60601-1-2:2014)
Interference may occur in the vicinity of equipment
marked with the following symbol:
Technical Data
a)
, should be less than
b)
.
29