TUBE CHECK WITH ROTOR CONTROL PART 2

▪ Switch ON the generator.
The system and the generator should be in ready condition.
Check of the tube vacuum:
▪ Push and hold PREP.
▪ Switch OFF the generator.
The anode is in free run now.
▪ Listen at the tube housing:
If the anode speed is down within 2 ... 3 minutes the tube is defective (no vacuum) and must be replaced.
If a decrease of anode speed is not audible within several minutes the vacuum is still OK.
▪ Switch the generator back ON.
A generator with high-speed rotor control brakes the tube automatically (chapter 8).
At generators with low -speed rotor control wait for ready condition, push PREP once and let go of the
PREP switch to brake the tube.