Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Tube adapted
Switch the number of exposures given for every data set of the table.
kV
10 seconds pause
145
1 minute pause
148
148
148
10 seconds pause
148
1 minute pause
150
150
150
10 seconds pause
150
Conv test Optimus
mA
ms
max mA
100
10
50
10
500
200
250
max mA
100
10
50
10
500
200
250
max mA
100
CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
exposures
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
ALL RIGHTS RESERVED
Tube not adapted
kV
mAs
10 seconds pause
145
100
1 minute pause
148
0.5
148
5
148
50
10 seconds pause
148
100
1 minute pause
150
0.5
150
5
150
50
10 seconds pause
150
100
39