Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
▪ Lead the primary power cable(s) of QC13:1 (front and rear) through the current transformer(s) at each converter
and reestablish the connection to the capacitor.
Optimus 65/80 with 2 converters:
Both current transformers must be inserted the same way.
(Polarity, secondary transformer pins are marked with 1 and 2).
▪ Reestablish primary cable(s) at QC3:1 too.
▪ Reestablish all connections (mains cables of converter(s) at ENK1 :1 :3 :5.