Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Optimus 65/80 tube not adapted:
(Fig. 19)
(Fig. 20)
If waveforms are not symmetrical replace HV transformer.
The primary current values must be within the range of the lower chart of page 28 ±10%.
12.1
M
EASURING OF THE RESONANCE FREQUENCY
The resonance frequency of
– the parallel primary coils (anode and cathode) at a 50kW generator driven by 1 converter only
or
Conv test Optimus