Philips Optimus Manual page 26

Converter test kit for optimus 50/65/80 generators release 3.x with converters 4512 104 7231x
Table of Contents

Advertisement

▪ Start with 40kV, 50mA, 10ms (= 0.5mAs). If the tube is not adapted set 40kV 0.5mAs.
▪ Establish the second side of the test cable with resistor at backpanel plug
EZX18 (Optimus 50)
or
EZX19 (OM 65/80) (backpanel layout page Z2-5.2).
If EZX19 is used take 1WA X42 or 2WA X42 or 1WB X42 (see layout Z2-15.1).
NOTE
▪ Switch on exposure.
▪ Compare the measured result with the four possible cases of the table.
Case 1:
Both, anode and cathode side, have the same kV rising slope and value up to the point when the exposure
terminates (which is almost the set kV).
If so, increase kV value in 10kV steps up to the programmed max kV tube limit or the max. value which is used
for application and check results with the table cases 1, 2 and 3 accordingly
(settings 50mA + 10ms or 0.5mAs for a non-adapted tube must remain!).
If the result is not case 1 carry out case dependent
exposure.
CAUTION
26
If the test cable with resistor is connected, mA-control EZ119 might display a FATAL
error condition (LED on) after a certain time. Then remove the plug and warmstart the
generator.
Do not put the plug back in before the generator is in ready condition. Then continue.
Case 4 must not be carried out with a value higher than 40kV.
CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
actions given in the table below and switch a second
Conv test Optimus