Converter test kit for optimus 50/65/80 generators release 3.x with converters 4512 104 7231x
Table of Contents

Advertisement

Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Optimus 65/80 tube adapted:
(Fig. 17)
(Fig. 18)
If waveforms are not symmetrical replace HV transformer.
The primary current values must be within the range of the lower chart ±10%.
32
CSIP Level 1 (08.0)
Conv test Optimus