Philips Optimus Manual page 49

Converter test kit for optimus 50/65/80 generators release 3.x with converters 4512 104 7231x
Table of Contents

Advertisement

Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Tube adapted
kV
140
140
140
10 seconds pause
140
1 minute pause
145
145
145
10 seconds pause
145
1 minute pause
148
148
148
10 seconds pause
148
1 minute pause
150
150
150
10 seconds pause
150
1 minute pause
Conv test Optimus
Exposure parameters for conditioning
mA
ms
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
# Exposures
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
ALL RIGHTS RESERVED
Tube not adapted
kV
mAs
140
0.5
140
5
140
50
10 seconds pause
140
100
1 minute pause
145
0.5
145
5
145
50
10 seconds pause
145
100
1 minute pause
148
0.5
148
5
148
50
10 seconds pause
148
100
1 minute pause
150
0.5
150
5
150
50
10 seconds pause
150
100
1 minute pause