Converter test kit for optimus 50/65/80 generators release 3.x with converters 4512 104 7231x
Table of Contents

Advertisement

Tube adapted
kV
80
80
80
10 seconds pause
80
1 minute pause
90
90
90
10 seconds pause
90
1 minute pause
100
100
100
10 seconds pause
100
1 minute pause
110
110
110
10 seconds pause
110
1 minute pause
120
120
120
10 seconds pause
120
1 minute pause
130
130
130
10 seconds pause
130
1 minute pause
48
Exposure parameters for conditioning
mA
ms
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
10
50
10
500
200
250
max. mA
100
CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
# Exposures
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
< 1 >
ALL RIGHTS RESERVED
Tube not adapted
kV
mAs
80
0.5
80
5
80
50
10 seconds pause
80
100
1 minute pause
90
0.5
90
5
90
50
10 seconds pause
90
100
1 minute pause
100
0.5
100
5
100
50
10 seconds pause
100
100
1 minute pause
110
0.5
110
5
110
50
10 seconds pause
110
100
1 minute pause
120
0.5
120
5
120
50
10 seconds pause
120
100
1 minute pause
130
0.5
130
5
130
50
10 seconds pause
130
100
1 minute pause
Conv test Optimus