Philips Optimus Manual page 44

Converter test kit for optimus 50/65/80 generators release 3.x with converters 4512 104 7231x
Table of Contents

Advertisement

▪ If the UCOMP value does not match the requirements type in another factor duty cycle value,
transmit the factor by clicking on "Apply" with the left mouse button and push the active RGDV button to get
the new value validated.
▪ Switch an exposure.
The values are measured in the stationary condition. The transient behavior at the beginning of the exposure
is not taken into account.
Result:
In standby the
given in table 1 or 2, refer to figure 29:
kV actual value +
Oscillogram of a perfect alignment for 141kV
(Fig. 29)
a) 141kV setting (65/80kW only)
▪ Read the mean value of
meets the required reference of +1V.
kV
mA
setpoint
setpoint
141kV
200mA
Table 1: Factor duty cycle, settings 141kV (150kV limit)
Example how to correct the factor duty cycle:
PCB kV_control 4:
▪ If the mean value of
▪ Check also the kV peak value AV HT (not the overshoot), it must be 138kV for 141kV setpoint
(see scope figure 30).
▪ Remove short link EZ130 X23 GAIN_IN.
▪ Record the results in table 1.
44
U
value is at about +11V, during exposure the mean value UCOMP must be as
COMP
UCOMP
U
for 141kV (see scope figure 29 or 30), correct the factor duty cycle till
COMP
U
PCB type
PCB kV_control 4:
U
is:
> +1.5V
COMP
< +0.5V
CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Tolerance
kV peak
COMP
of
+1V
±0.5V
increase the factor duty cycle in steps of 0.01
decrease the factor duty cycle in steps of 0.01
ALL RIGHTS RESERVED
UCOMP
UCOMP for 141kV
Factor duty cy‐
AV HT
cle:
138kV
Conv test Optimus
U
COMP
Date