Overview Of Error Codes - LINET Sprint 100 Service Manual

Universal transport stretcher with scales
Table of Contents

Advertisement

4.1.2 Overview of error codes

Letter indicating the type of error
A
B
F
Keyboard error code
(letter A)
0x1
0x2
0x4
0x8
0x10
0x20
0x100
0x200
0x400
0x800
0x1000
0x2000
Scale system hardware error code
(letter B)
0x1
0x2
0x4
0x8
0x10
0x20
0x80
0x2000
0x4000
D9S001ES1-0101
Type of error
Keyboard errors
Scale system hardware errors
Weighing system errors
1
2
3
Meaning
Long press (≥ 60) detected on the ZERO/T button (button No. 1)
Long press (≥ 60) detected on the WEIGHT/CLEAR button (button No. 2)
Long press (≥ 60) detected on the kg lb button (button No. 3)
Long press (≥ 60) detected on the Bed Exit Alarm ON/OFF button (button No. 4)
Long press (≥ 60) detected on the Bed Exit Alarm pause button (button no. 5)
Long press (≥ 60) detected on the Bed Exit Alarm mode selection button (button No. 6)
Error detected on ZERO/T button (button No. 1)
Error detected on WEIGHT/CLEAR button (button No.2)
Error detected on kg/lb button (button No. 3)
Error detected on the Bed Exit Alarm ON/OFF button (button No. 4)
Error detected on the Bed Exit Alarm pause button (button No. 5)
Error detected on the Bed Exit Alarm mode selection button (button No. 6)
Meaning
Overvoltage on the scale system processor
Undervoltage on the scale system processor
Voltage measurement error on the scale system processor
Overheating of the scale system processor
The processor temperature of the scale system is below the permitted limit
Error of temperature measurement of the scale module processor
Other hardware error
Error of Bed Exit Alarm function
Button measurement error (control measurement mismatch)
4
5
6
23