LINET ELEGANZA 1 Instructions For Use And Technical Description

LINET ELEGANZA 1 Instructions For Use And Technical Description

Hospital bed for standard care
Table of Contents

Advertisement

Instructions for Use
and Technical Description
ELEGANZA 1
Hospital bed for standard care
D9U001GT0-0101
Version: 09
Publication Date: 2020-02

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ELEGANZA 1 and is the answer not in the manual?

Questions and answers

Dimitar Dimitrov
February 25, 2025

Hi, writing from Hospital in Bulgaria. We have beds, type Linet Eleganza 1, one of them have a problem. Reepedly sound, silence and again. I send video. I ask for suport, at this problem. Thanks.

1 comments:
Mr. Anderson
February 25, 2025

The repeated beeping sound in the LINET Eleganza 1 bed, lasting 1.1 seconds on and 1.1 seconds off for 3 minutes, is associated with the brake signal. This issue occurs only in versions equipped with a brake signal. It indicates that the brake may not be properly engaged or there is a malfunction in the brake system. Checking the brake mechanism and ensuring it is correctly engaged may resolve the issue.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for LINET ELEGANZA 1

  • Page 1 Instructions for Use and Technical Description ELEGANZA 1 Hospital bed for standard care D9U001GT0-0101 Version: 09 Publication Date: 2020-02...
  • Page 2 D9U001GT0-0101 Version: 09 Publication Date: 2020-02 Copyright © LINET, s.r.o., 2020 Translation © LINET, 2020 All rights reserved. All trademarks and brands are the property of the appropriate owners. The manufacturer reserves the right to changes in the contents of the instructions for use that relate to the product´s technical regulations.
  • Page 3: Table Of Contents

    12.8.3 Split Plastic Siderails (SR22) .........53 5 Product Description .............18 12.8.4 Full length siderails ..........53 5.1 Eleganza 1 (1GTP) with Single Collapsible Siderails . 18 12.8.5 Split Plastic Siderails (SR46) .........54 5.2 Eleganza 1 (1GTP) with Split Plastic Siderails (SR22) 12.9 Castor Control and Bed Transport ......
  • Page 4: Symbols And Definitions

    19.2.2 Outside Europe ............69 20 Warranty ..............70 21 Standards and Regulations ........70 1 Symbols and Definitions 1.1 Warning Notices 1.1.1 Types of Warning Notices Warning notices are differentiated by the type of danger using the following key words: ► CAUTION warns about the risk of material damage. ►...
  • Page 5: Symbols On The Package

    1.4 Symbols on the Package FRAGILE, HANDLE WITH CARE THIS WAY UP KEEP DRY (PROTECT FROM HUMIDITY) PAPER RECYCLING SYMBOL DO NOT USE HAND TRUCK HERE DO NOT STACK DURING STORAGE D9U001GT0-0101_09...
  • Page 6: Symbols And Labels On The Bed

    1.5 Symbols and Labels on the Bed READ INSTRUCTIONS FOR USE WARNING THERMAL PROTECTION OF TRANSFORMER ONLY SUITABLE FOR INDOOR USE PROTECTION AGAINST ACCIDENTS DUE TO ELECTRICAL CURRENT – TYPE B APPLIED PARTS SAFETY ISOLATING TRANSFORMER, GENERAL CE MARK OF CONFORMITY WITH EU REGULATION SAFETY MARK –...
  • Page 7 USE MATTRESS RECOMMENDED BY MANUFACTURER WEIGHT OF BED DO NOT PUT ANY OBJECTS HERE DESIGNATION OF HOSPITAL BED FOR ADULTS ALTERNATING CURRENT DIRECT CURRENT EAC MARKING WEEE SYMBOL (RECYCLE AS ELECTRONIC WASTE, DO NOT PUT INTO THE HOUSEHOLD WASTE) MEDICAL DEVICE (compatible with Medical Device Regulation) RECYCLING SYMBOL DO NOT POLLUTE THE ENVIRONMENT D9U001GT0-0101_09...
  • Page 8 HEAD BOARD UNLOCKED OR FOOT BOARD UNLOCKED HEAD BOARD LOCKED OR FOOT BOARD LOCKED D9U001GT0-0101_09...
  • Page 9: Serial Label With Udi

    1.6 Serial Label with UDI Pictures of serial labels below serve just for explanation of the signs and fi elds on the serial labels. Fig. Serial Label with UDI (Eleganza 1 - version 1) Fig. Serial Label with UDI (Eleganza 1 - version 2)
  • Page 10: Acoustic Signalisation (Eleganza 1 With Pb43 Control Unit)

    REPEATED BEEP during 3 minutes: 1,1s sound / 1,1s silen- Brake Signal (only version with Brake Signal) 1.8 Acoustic signalisation (Eleganza 1 with PB11 control unit) There are no acoustic signals in the case of Eleganza 1 bed with PB11 control unit. D9U001GT0-0101_09...
  • Page 11: Visual Signalisation

    1.9 Visual signalisation 1.9.1 Mains Power LED (Attendant Control Panel) MAINS POWER LED MEANING connected to the mains flashing: 0,6s lit / 0,6s not lit keyboard error (flashing inverted to Lock LED) error (first fault) flashing: 0,1s lit / 0,1s not lit service mode not lit disconnected from the mains power transformer switching error 1.9.2 Accumulator Indicator (Attendant Control Panel, MiniACP) ACCUMULATOR INDICATOR MEANING...
  • Page 12: Definitions

    1.10 Definitions Basic Bed Configuration the pricelist model configuration, not including a mattress Bed Weight The value depends on the product configuration, accessories or customer adjustments. Clearance of Undercarriage the height from the floor to the lowest point of the undercarriage between the castors, for the manipulation of accessories under a braked bed in the standard position Duty Cycle...
  • Page 13: Abbreviations

    1.11 Abbreviations AC ( ~ ) Alternating Current Attendant Control Panel European Conformity Cardiopulmonary Resuscitation Sound Intensity Unit DC ( Direct Current Configuration number Electromagnetic Compatibility Field-effect transistor High Frequency High Pressure Laminate Hardware Intensive Care Unit INT. Duty Cycle Ingress Protection Intravenous Light Emitting Diodes...
  • Page 14: Safety Instructions

    2 Safety Instructions WARNING! Eleganza 1 bed should be left in its lowest position when the patient is unattended in order to reduce risk of injury due to falls! WARNING! Siderails of Eleganza 1 should be located in the „up“ position to reduce the risk of the patient accidentally...
  • Page 15 WARNING! Incompatible mattresses can create hazards. WARNING! The bed is intended for adults. ►Follow the chapter Intended use. WARNING! Length adjustment of the bed must be proportional to the height of patient! WARNING! Risk of trapping or squeezing because of patient´s body constitution disproportionate to the size of mattress support platform! WARNING! Risk of injury or death due using of incorrect equipment!
  • Page 16 ► Follow the instructions for use carefully. ► Any non-observance of these instructions for use may lead to injuries or damage of property. ► Exclusively use the bed if it is in perfect working order. ► If necessary, check the bed functions daily or at each shift change. ►...
  • Page 17: Additional Safety Instructions For Extra Safe Bed

    LINET®’s efforts in research, design and manufacture ensure LINET® products are of the highest quality and fit for their intended purpose. However, LINET® can take no responsibility for any damage to the products or any harm to patients, staff or other indivi- duals resulting from: ■   ...
  • Page 18: Product Description

    Single collapsible siderails Head Board Handset Castor Fifth Castor CPR lever (Backrest release) Central castor control lever Calfrest Handle Corner Bumper Linen shelf MiniACP (Supervisor board) NOTE For safe, easy handling, LINET® recommends having two technicians assemble the bed. D9U001GT0-0101_09...
  • Page 19: Eleganza 1 (1Gtp) With Split Plastic Siderails (Sr22)

    Split collapsible siderails Control element, integrated in siderail Head Board Castors Siderail release mechanism (based on tipping down) Central castor system lever Linen shelf Supervisor board NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service. D9U001GT0-0101_09...
  • Page 20: Eleganza 1 (1Gtp) With Split Plastic Siderails (Sr46)

    Mattress Platform (Removable Plastic Parts) Foot Board Linen Shelf Brake Lever (optional) Siderail Release Mechanism (based on lifting towards the user) Castor Brake Lever (optional) Corner Bumper NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service. D9U001GT0-0101_09...
  • Page 21: Eleganza 1 (1Gtd) With Wooden Head Board And Foot Board

    Foot board Siderail release mechanism Mattress base Head board Castors Control element, integrated in siderail Single collapsible siderail Central castor system lever Linen shelf NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service. D9U001GT0-0101_09...
  • Page 22: Eleganza 1 (1Gtp) - Single Collapsible Siderails Controlled From Top

    Castor 150 mm Lockable box for mains cable Accessory rail CPR lever Central castor system lever Linen shelf with non-removable Attendant Control Panel Segufix holders NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service. D9U001GT0-0101_09...
  • Page 23: Eleganza 1 (1Gtp) - Single Collapsible Siderails Controlled From Bottom

    Bed foot with the possibility of fixation to the floor Lockable box for mains cable Accessory rail Siderail release mechanism CPR lever Linen shelf with non-removable Attendant Control Panel Segufix holders NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service. D9U001GT0-0101_09...
  • Page 24: Technical Specification

    6 Technical Specification 6.1 Type B Applied Parts All the accessories the patient can reach are Type B Applied Parts. List of type B applied parts: ■    Handset, Additional Attendant Control Panel, Control Element integrated in Siderail ■    Siderails ■    Head board and foot board ■   ...
  • Page 25: Mechanical Specification (Shortened Version)

    6.3  Mechanical Specification (shortened version) Parameter Value Dimensions (without Bed Extension) 218 cm (205 cm) x 99,5 cm (depending on confi- guration) Siderail lenght Split siderail ■    Foot siderail 96,5 cm ■    Head siderail 49,5 cm Collapsible siderails 146,5 cm Siderail height Split siderail ■    Foot siderail 35 cm ■   ...
  • Page 26: Environment Conditions

    Sound Pressure Level 60 dB(A) NOTE Upon request, LINET® can deliver hospital beds with electrical specifications that comply with regional standards (custom voltage, different mains plugs). Electrical system of Eleganza 1 is characterized by overcurrent protection against overload of each actuator.
  • Page 27: Manufacturer Instructions - Electromagnetic Emissions

    List of used cables: Mains cable, maximum length 6 m ACP Attendant Control Panel, maximum length 3m Handset, maximum length 3m WARNING! Use of the accessories, converters and other cables, than specified and provided by manufacturer of this bed  could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this bed and lead to improper operation.
  • Page 28: Use And Storage Condition

    Table 1 - IMMUNITY to RF wireless communications equipment Test frequency (MHz) Band (MHz) Service Modulation Immunity Test Level 380 - 390 TETRA 400 Pulse modulation 18 Hz 430 - 470 GMRS 460, FRS 460 FM ± 5 kHz deviation 1 kHz sine 704 - 787 LTE band 13, 17...
  • Page 29: Scope Of Delivery And Bed Variants

    □ Segufix holder ■ Powder Coating Colours □ powder-coated metal parts, RAL 9002 (white) □ powder-coated metal parts, RAL 9006 (light grey) 8.2.2 Eleganza 1 Model 1GTP Extra Safe Bed s = standard o = optional ■ Mattress Platform □ mattress platform consisting of fixed HPL parts (s) ■...
  • Page 30: Eleganza 1 Model 1Gtd Nursing Bed

    □ 4 head and foot end bushings for accessories (s) ■ Powder Coating Colors □ RAL 9002 (grey) (s) □ Powder-coated metal parts, RAL 9006 (light grey) + RAL 7043 (dark grey) (o) 8.2.3 Eleganza 1 Model 1GTD Nursing Bed s = standard o = optional ■ Mattress Platform □...
  • Page 31: Putting Into Service

    9 Putting into Service WARNING! Risk of injury when working on the bed! ► Ensure that the bed is disconnected from the mains connection prior to putting into service, putting out of service and maintenance. ► Ensure that the castors are locked prior to putting into service, putting out of service and maintenance. CAUTION! Material damage due to incorrect putting into service! ►...
  • Page 32: Accumulator Activation

    9.1 Accumulator Activation 9.1.1 Placement of Control Section 9.1.2 Removing the Isolating Foil 9.1.3 Isolating Foil Check if isolating foil is complete and undamaged as shown: If isolating foil is damaged, contact the manufacturer’s service department immediately. NOTE: It is recommended to wear gloves when removing the isolating foil. D9U001GT0-0101_09...
  • Page 33: Head Board And Foot Board

    9.2 Head Board and Foot Board WARNING! Risk of injury when inserting the head board and foot board! ► To insert head board and foot board into corner posts, hold them by the corner handles on top with both hands. ►...
  • Page 34 Locked Unlocked Install head board and foot board as follows: ► Unlock safety levers on corner posts. ► Slide head board and foot board into slots on corners posts with coloured panels outside. ► Lock safety levers on corner posts. Remove head board and foot board as follows: ►...
  • Page 35: Mattress Support Platform

    9.3 Mattress support platform Eleganza 1 bed has 4-part Mattress support platform consisting of Backrest, Seat section, Thighrest and Calfrest. The Mattress support platform has 7 removable plastic sections of Mattress support platform (1, 2, 3, 4, 5, 6 and 7).
  • Page 36: Shortened Bed

    Use the bed shortening carefully when a patient is on the bed to avoid risk of crushing, trapping or squeezing! Eleganza 1 medical bed can be delivered in the version with bed shortening. Purpose of this shortened version is to facilitate bed transport.
  • Page 37: Potential Equalisation

    9.4 Potential Equalisation The bed is equipped with a standard protective connector. This connector is used for potential equalisation between the bed and any intravascular or intracardiac device connected to the patient to protect the patient from static electric shocks. Fig.
  • Page 38: Before Use

    9.5 Before Use CAUTION! Material damage due to temperature difference! ► If there is a considerable temperature difference between the bed and the place of operation (after transport/storage), leave bed unconnected for 24 hours to allow the temperature to equalise. Prepare the bed for service as follows: ►...
  • Page 39: Power Cable

    10 Power Cable CAUTION! Disconnecting bed from the mains does not stop motions of the bed! ► Stop the bed before disconnection bed from the mains. Attachment plug is means of connecting and disconnecting bed from the mains. Power supply cable (mains power cable) must be attached with a hook at the head end of the bed during transport. Where the integrity of the external protective conductor in the installation or its arrangement is in doubt ►...
  • Page 40: Accumulator

    11 Accumulator CAUTION! Risk of reducing accumulator durability due to incorrect use! ► Use bed on accumulator only in crisis situations (e.g.: power blackout, patient complications during transport, etc.). ► After reconnecting bed to the mains charge accumulator to full capacity (see chart Accumulator charge status). CAUTION! Risk of damage or destruction of accumulator! ►...
  • Page 41: Status Faulty Accumulator

    The LED on ACP indicates the accumulator charge status: Yellow LED Accumulator charge status Not lit Accumulator capacity is sufficient (charging completed) Short flashing (short, intermittent illumination) Accumulator is charging - continue charging until the LED is (circa 1.8 sec.) extinguished.
  • Page 42: Removing The Bed From Service

    11.3 Removing the Bed from Service How to remove the bed from service: ► Disconnect the bed from the mains. ► Disconnect the ground wire. ► Deactivate the accumulator (see Deactivating the Accumulator). ► Remove accessories. To prevent damage during storage: ►...
  • Page 43: Manipulation

    12 Manipulation WARNING! Risk of injury when adjusting the bed! ► Ensure there are no body parts between the mattress platform elements and the mattress platform frame when adjusting the bed. ► Ensure there are no body parts below the mattress platform frame before adjusting the bed. ►...
  • Page 44: Acp - Attendant Control Panel

    12.2 ACP – Attendant Control Panel ACP Attendant Control Panel is main control panel of the bed. Nurse controller can be placed in the shelf. We recommend placing ACP panel on the head board or foot board or hold it in the hands while controlling the bed. Fig.
  • Page 45: Activating Go Button

    12.2.2 Activating GO Button The GO button 17 activates the keypads on all control elements except the foot control. A GO button is included on a number of different control elements. The function of the GO button is identical on all control elements. After pressing GO button 17, the keypad will remain active for 3 minutes.
  • Page 46: Handset

    If required, prevent the patient from adjusting the bed as follows: ► Disable functions. ► Unplug handset (if the bed is equipped with Plug and Play). NOTE: The handset can be plugged into another LINET® hospital bed with the Plug and Play slot. D9U001GT0-0101_09...
  • Page 47: Satellite Control Panel

    12.4 Satellite Control Panel The satellite control panel is an optional control element. The satellite control panel is attached to the backrest with a flexible arm. 1. Activating GO Button 2. Thighrest Adjustment Buttons 3. Backrest Adjustment Buttons 4. Height Adjustment Buttons Fig.
  • Page 48: Control Element Integrated In Siderail

    12.5 Control Element Integrated in Siderail It is possible to equip the Eleganza 1 bed with integrated illuminated controllers in the siderails. The optional control element is attached to the siderail. The controllers are positioned on both sides of the bed.
  • Page 49: Foot Control

    12.6 Foot Control WARNING! Risk of injury due to unlocked foot control! ► Ensure that the foot control is locked if it is not in use. ► Ensure no body parts are trapped between moving parts of bed and mattress platform. ►...
  • Page 50: Cpr Backrest Release

    12.7 CPR backrest release WARNING! Risk of injury due to lowering the backrest too quickly! ► Ensure single collapsible siderails are in their lowest position. ► Ensure split plastic siderails are in their highest position. ► Ensure there are no body parts between siderails and backrest. ►...
  • Page 51: Siderails

    12.8 Siderails WARNING! Risk of injury due to crushing! ► Ensure that there are no objects or body parts between siderail and mattress platform when folding the siderail up or down. ► Ensure that no body part or object are under the siderail when folding the siderail down. ►...
  • Page 52: Single Collapsible Siderails - Controlled From The Bottom

    Fig. Release Mechanism (inner side of collapsible siderail) To fold siderail down: ► Take release mechanism and push siderail towards head end (1). ► Press button placed on the inner side of the siderail (2) to unlock the siderail. ► Fold siderail down as required (3).
  • Page 53: Split Plastic Siderails (Sr22)

    12.8.3 Split Plastic Siderails (SR22) WARNING! Risk of injury due to patient falling out of bed! ► Ensure that folded-up siderails are securely anchored. ► Press against siderails from inside in order to check stability. ► Hospital personnel are responsible for ensuring siderails are folded-up when the bed is occupied. The split plastic siderails are components of the bed, the siderails cannot be removed.
  • Page 54: Split Plastic Siderails (Sr46)

    12.8.5 Split Plastic Siderails (SR46) WARNING! Risk of injury due to patient falling out of bed! ► Ensure that folded-up siderails are securely anchored. ► Press against siderails from inside in order to check stability. ► Hospital personnel are responsible for ensuring siderails are folded-up when the bed is occupied. The split plastic siderails are components of the bed, the siderails cannot be removed.
  • Page 55: Castor Control And Bed Transport

    12.9 Castor Control and Bed Transport CAUTION! Material damage due to incorrect transport or involuntary movement! ► Prior to assembly, disassembly and maintenance, ensure the castors are locked. ► Ensure the castors are locked while the bed is occupied and/or not being transported. ►...
  • Page 56: Individual Braking Control

    Simple Castor Control Levers (optional): Fig. Simple Castor Control Lever Castor control with central braking system: Braked: All of the castors are locked. Unrestricted Movement: All of the castors are unlocked. Forward Movement: The front left castor is locked. The bed moves straight ahead. If the bed is equipped with a fifth cas- tor, this castor determines the direction of movement.
  • Page 57: Equipment

    3 minutes after pressing of the GO button and it is not lit in the other cases. 14 Mattress The manufacturer recommends the use of the following mattress systems on the Eleganza 1 bed: ►...
  • Page 58: Accessories

    Risk of injury due to damaged accessories! ► Use exclusively accessories in perfect condition. NOTE: The manufacturer assumes no responsibility for the use of accessories not approved by the manufacturer. NOTE: All accessories conform to IEC 60601-2-52. The following accessories are suitable for the Eleganza 1: ■ Lifting pole □ powder-coated □...
  • Page 59: Lifting Pole

    A plastic grab handle with an adjustable strap is attached to the lifting pole. NOTE: The date of manufacture is marked on the grab handle. NOTE: LINET® recommends replacing the plastic grab handle every four years. 15.2 Infusion Stands WARNING! Risk of injury due to use of unsuitable accessories! ►...
  • Page 60: Name Holder

    15.3 Name holder Name holder is used for placing patient’s name on it. Name holder must be placed under the head board or foot board handles so it will not collide with movable parts of the bed. This must be verified after installation of name holder. 15.4 Urinary bottle holder WARNING! Risk of collision when folding down the siderails!
  • Page 61: Siderail Extension - „Extender®" For Siderails Controlled From Top (Sr55)

    Eleganza Protector® is not a component of the bed. Eleganza Protector® is optionally available Fig. Closed Eleganza Protector® Fig. Opened Eleganza Protector® Mount Eleganza Protector® to closed position as follows: ► Put the tube of protector (2) into bushing near foot end of the bed (3) so the protector is facing the siderail. ►...
  • Page 62: Siderail Extension - „Extender®" For Siderails Controlled From Bottom (Sr57)

    Fig. Installation of Extender® Fig. Securing Extender® Install the Extender® as follows: ► Place the Extender® (without securing screws) on the upper edge of the siderail. Siderails must be in upper position. Square hole is pointed into the mattress platform. ►...
  • Page 63 It is recommended to use Extender® with the following mattresses: ■ Passive mattress: higher than 16 cm, with maximum height 26,5 cm ■ Active mattress: with maximum height 26,5 cm ● Virtuoso Fig. Installation of Extender® Fig. Securing Extender® Install the Extender® as follows: ►...
  • Page 64: Cleaning/Disinfection

    16 Cleaning/Disinfection WARNING! Risk of injury when working on the bed! ► Prior to assembly, disassembly, cleaning and maintenance, ensure that all adjustment functions are locked. ► Ensure the bed is disconnected from the mains during cleaning process. ► Pay extra attention when cleaning any movable or controlling mechanisms of the bed to prevent involuntary activation, entrapping or crushing.
  • Page 65: Safety Instructions For Cleaning And Disinfection Of The Bed

    16.1 Safety Instructions for Cleaning and Disinfection of the Bed Preparation for cleaning: ► Drive the bed on a place where the cleaning process will be performed and then brake the bed. ► Position the mattress platform to its highest positions and also position the backrest and thighrest parts so the back side of those parts are accessible for cleaning.
  • Page 66: Modes Of Cleaning And Disinfection

    16.3 Modes of Cleaning and Disinfection Part of bed – Eleganza 1 Daily C&D Changing patient C&D Complete C&D Siderails (stickers): ● Split plastic    ● Single collapsible    Siderails mechanism    Head board and foot ...
  • Page 67: Troubleshooting

    17 Troubleshooting DANGER! Danger to life due to electric shock! ► If a fault occurs ensure the electric motor, power box and other electrical parts checked by qualified personnel only. ► Do not open protective covers of the electric motor or power box. Error/Fault Cause Solution...
  • Page 68: Maintenance

    Ensure that maintenance is performed exclusively by manufacturer´s customer service or by authorised service personnel certified by the manufacturer. ► If the defect cannot be repaired, do not use the bed. LINET ® recommends attaching the maintenance plaque to the bed. 18.1 Regular maintenance ► Check regularly movable parts for wear.
  • Page 69: Disposal

    Based on the Directive No. 2002/96/ EC (Directive WEEE - Waste, Electric and Electronic Equipments) the company LINET, s. r. o. is registered in the List of Electric and Electronic Equipment Producers (Seznam výrobců elektrozařízení) on the Ministry of the Environment of the Czech Republic...
  • Page 70 20 Warranty LINET ® will only be held responsible for the safety and reliability of products that are regularly serviced, maintained and used in accordance with the safety guidelines. Should a serious defect arise that cannot be repaired during maintenance: ►...