4.4

Inspection of the light cutoff function

VISERA XENON LIGHT SOURCE CLV-S40
Do not look directly into the distal end of the endoscope or
the output connector of the light source while they are
emitting light. Eye damage may result.
If you observe conspicuous light emission from the output
connector after disconnecting the endoscope, stop using the
light source and contact Olympus.
Confirm that no light is emitted from the light source by following the procedures
described below.
1.
While emitting light from the endoscope, disconnect the endoscope. Place a
piece of paper over the output connector of the light source and confirm that
no light is being emitted from it (see Figure 4.6).
2.
Reconnect the endoscope to the output connector and confirm that light is
emitted from the distal end of the endoscope (see Figure 4.5).
Checking side.
Figure 4.6
Paper
Chapter 4 Inspection
39