Page 1
Confluency checker NOTE: This instruction manual is for the Olympus confluency checker CKX-CCSW. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this software, we recommend that you study this manual thoroughly before using the software, and always keep this manual at hand when using the software.
Contents Introduction ............................... 1 Handling Precautions ..........................2 Cautions in use ................................2 System compatibility ..............................2 PC and software ................................3 1 System diagram ............................4 2 Installation ..............................5 2-1 Installation................................5 2-2 Activation .................................. 8 2-3 Uninstallation ................................ 8 3 Basic operations .............................
Page 3
CKX-CCSW 5 Screen flow ............................... 38 5-1 [Main] screen ..............................38 5-2 Displaying the submenu .......................... 39 5-3 Displaying the [Setting] screen ......................39 6 Various settings ............................. 40 6-1 [Main] screen ..............................40 Setting the source ..................................40 6-2 Submenu ................................41 Setting the area of culture vessel ..........................41...
Page 4
7 Effective measurement of cell count/confluency ........... 53 7-1 Zoom area ................................53 7-2 Using the cell scale bar effectively ....................53 Setting the cell scale bar ..............................54 8 Specifications ............................55 9 Troubleshooting ............................ 56...
CKX-CCSW Introduction The confluency checker CKX-CCSW is a product to measure the cell count and the confluency from the image where the culture cells are acquired with the Olympus microscope digital camera DP22/DP27 or other cameras. Contents in this instruction manual This instruction manual covers only the contents concerning the confluency checker CKX-CCSW.
Cautions in use Image data 1. The recorded image data may be lost (destroyed) in the following cases. Olympus is not liable for the loss (destruction) of the recorded data. • When the user or a third party carried out a repair to the product.
PCs whose performances in combination use with this system were confirmed. 1. The PC data may be lost (corrupted) unexpectedly, so make frequent backups of the data you acquired. Olympus is not liable for any damage due to the use or disability in use of this product, including compensation for lost data.
} Consult Olympus for the compatible microscope and camera adapter. The camera head may not work properly in combination with some PCs. Refer to Olympus website to check the PCs whose performances in combination use with this system were confirmed.
(For the change of user account, refer to the Help of Windows.) 2-1 Installation The procedures to install this product is described below. Open (decompress) the file downloaded from the Olympus web site on the desktop screen of PC where this product will be installed. Click "Setup.exe".
Page 10
Select the folder to install the software and click [Next]. The following folder is specified as the initial setting. 32-bit version C:\Program Files\Olympus\Micro\CKX-CCSW 64-bit version C:\Program Files (x86)\Olympus\Micro\CKX-CCSW Click [Install]. The installation starts. During the installation process, if "Would you like to install this device software?] regarding the...
Page 11
CKX-CCSW Click [Finish]. This is the end of the installation.
2-2 Activation If passing 30 days of more after the date of installation, you must activate this product using the serial code. Without activation, you will not be able to use this software. The procedures to activate this product is described below. Click on the desktop.
· Windows 7 Click on the desktop screen. Or click in the order of [Start] button - [All programs] - [Olympus] - [CKX-CCSW]. If the error or the warning screen appears after the software starts, click [OK]. · Windows 8/8.1 Display the desktop screen and click Or on the start screen, click [CKX-CCSW.exe] in [All applications].
Button to minimize Minimizes the screen. Button to close Closes CKX-CCSW * These fields may not be displayed depending on the camera to be used and software settings. Exiting the software Exit this product according to the procedures described below.
CKX-CCSW Initial settings This section describes the initial setting procedures of this product. This operation is necessary when you start this product for the first time. After the second time, the product starts in the status reflecting the last settings when exiting the software.
Page 16
Click [Main] d . The [Main] screen appears. In [Source], click e and select the source. Select the source from followings. · [DP]: Appears when DP22/DP27 is connected. (Keep this status when measuring while acquiring the image with DP22/ DP27.) ·...
CKX-CCSW 3-2 Measuring while acquiring the image with DP22/DP27 You can measure the cell count and the confluency in the culture vessel while acquiring the image with DP22/DP27. Measuring the cell count in the culture vessel while acquiring the image with DP22/DP27 Start this product according to the procedures described on page 9.
Page 18
If the submenu is not opened, click [Configuration] b to open the submenu. Click c of [Area of Culture Vessel] to display the items, and select the area of the culture vessel to be used. Items to be displayed are described below. If [Custom] is selected, input the numerical value in the d field.
Page 19
CKX-CCSW While viewing the cell image displayed in the preview area, adjust the threshold on the [Cell Count Setting] screen so that the point (shown in red as default) is displayed one by one in the cell. There are two ways to adjust the threshold: one is to click h or drag i to adjust the slider and the other is to input the numerical value into j directly.
Page 20
Bring the cell into focus again according to the procedure in Click [Snap] o on the [Measurement Result View] screen. The image is acquired and the measurement results are displayed in p . Move the culture vessel with your hand to change the acquisition position, and repeat Each time you acquire the image, the value displayed in p changes.
CKX-CCSW Measuring the confluency in the culture vessel while acquiring the image with DP22/DP27 Start this product according to the procedures described on page 9. The [Main] screen appears in the control area. Click a in [Source] and select [DP].
Page 22
If the submenu is not opened, click [Configuration] b to open the submenu. Click c of [Area of Culture Vessel] to display the items, and select the area of the culture vessel to be used. Items to be displayed are described below. If [Custom] is selected, input the numerical value in the d field.
Page 23
CKX-CCSW Click [OK] h on the [Step 1] screen and drag the cell part in the image displayed on the preview area. The dragged line (shown in blue as default) appears and the [Step 2] screen is displayed. Click [OK] i on the [Step 2] screen and drag the non-cell part (background part) in the image displayed on the preview area.
Page 24
On the [Main] screen, set the number of images to be measured while acquiring. In [Number of Images], input the numerical value into the m field directly or click [+] or [-] in the n field. The more the number of images is increased, the higher the measurement accuracy becomes, but it takes longer time.
CKX-CCSW 3-3 Measuring from the acquired image You can measure the cell count and the confluency from the acquired image. Measuring the cell count from the acquired image Start this product according to the procedures described on page 9. The [Main] screen appears in the control area.
Page 26
If the submenu is not opened, click [Configuration] f to open the submenu. Click g of [Area of Culture Vessel] to display the items, and select the area of culture vessel used for acquiring the image. Items to be displayed are described below. If [Custom] is selected, input the numerical value in the h field.
Page 27
CKX-CCSW While viewing the cell image displayed in the preview area, adjust the threshold on the [Cell Count Setting] screen so that the point (shown in red as default) is displayed one by one in the cell. There are two ways to adjust the threshold: one is to click l or drag m to adjust the slider and the other is to input the numerical value into n directly.
Measuring the confluency from the acquired image Start this product according to the procedures described on page 9. The [Main] screen appears in the control area. Click [Setting] a . The [Setting] screen appears. Click [General] b . Click c . The screen to select the folder appears. Select the folder where the image to be used for measurement is saved and click [Select Folder].
Page 29
CKX-CCSW If the submenu is not opened, click [Configuration] f to open the submenu. Click g of [Area of Culture Vessel] to display the items, and select the area of culture vessel used for acquiring the image. Items to be displayed are described below.
Page 30
Click [OK] l on the [Step 1] screen and drag the cell part in the image displayed on the preview area. The dragged line (shown in blue as default) appears and the [Step 2] screen is displayed. Click [OK] m on the [Step 2] screen and drag the non-cell part (background part) in the image displayed on the preview area.
Page 31
CKX-CCSW When measuring multiple images, select all images to be used for measurement from images displayed in the thumbnail area. Clicking the image in the thumbnail area will change between the selected status (framed in blue) and the unselected status (not framed in blue).
DP27 3-4 Measuring while acquiring the image with cameras other than DP22/ You can measure the cell count and the confluency in the culture vessel while acquiring the image with cameras other than DP22/DP27. Measuring the cell count in the culture vessel while acquiring the image with cameras other than DP22/DP27 Start this product according to the procedures described on page 9.
Page 33
CKX-CCSW If the submenu is not opened, click [Configuration] b to open the submenu. Click c of [Area of Culture Vessel] to display the items, and select the area of the culture vessel to be used. Items to be displayed are described below.
Page 34
While viewing the cell image displayed in the preview area, adjust the threshold on the [Cell Count Setting] screen so that the point (shown in red as default) is displayed one by one in the cell. There are two ways to adjust the threshold: one is to click h or drag i to adjust the slider and the other is to input the numerical value into j directly.
Page 35
CKX-CCSW Move the culture vessel with your hand to change the acquisition position and bring the cell into focus according to the procedure in . Then, acquire the image of the specimen according to the operating procedures of the camera in use. Repeat this operation.
Measuring the confluency in the culture vessel while acquiring the image with cameras other than DP22/DP27 Start this product according to the procedures described on page 9. The [Main] screen appears in the control area. Click a in [Source] and select [Folder]. Place the culture vessel of the specimen to be measured on the stage of the microscope.
Page 37
CKX-CCSW If the submenu is not opened, click [Configuration] b to open the submenu. Click c of [Area of Culture Vessel] to display the items, and select the area of the culture vessel to be used. Items to be displayed are described below.
Page 38
Click [OK] h on the [Step 1] screen and drag the cell part in the image displayed on the preview area. The dragged line (shown in blue as default) appears and the [Step 2] screen is displayed. Click [OK] i on the [Step 2] screen and drag the non-cell part (background part) in the image displayed on the preview area.
Page 39
CKX-CCSW On the [Main] screen, set the number of images to be measured while acquiring. In [Number of Images], input the numerical value into the m field directly or click [+] or [-] in the n field. The more the number of images is increased, the higher the measurement accuracy becomes, but it takes longer time.
3-5 Measuring the cell count and the confluency at the same time You can measure the cell count and the confluency at the same time by ticking both [Count] and [Confluency] on the [Main] screen in the control area. Refer to the respective operating procedures for details. 3-6 Re-using the measurement settings When you measure other culture cells under the same conditions, you don't need to set [Area of Culture Vessel] or [Objective Magnification] specified during measurement and to calibrate [Count] or [Confluency] again.
・ Always use the interface cable (USB 3.0 cable) provided by Olympus. Using commercially available USB 3.0 cables or hubs does not guarantee the operation of the equipment. ・ Keep the interface cables well away from the equipment generating heat, such as the lamp housing of microscope.
CKX-CCSW 5-2 Displaying the submenu This section describes the procedure to display the submenu of the [Main] screen. Display the [Main] screen. Click [Configuration] a . The submenu is displayed. Clicking a again closes the submenu. 5-3 Displaying the [Setting] screen This section describes the procedure to display the [Setting] screen.
Various settings 6-1 [Main] screen This section describes the settings available on the [Main] screen. You can display the [Main] screen by clicking the [Main] tab a in the control area. Minimize Save Save destination Exp. time Select image source Source Configuration Area of...
CKX-CCSW 6-2 Submenu Setting the area of culture vessel This section describes the procedures to set the area of culture vessel of the cell. Display the [Main] screen. Click a , and select from the items displayed. Items to be displayed are described below.
Setting the adapter magnification This section describes the procedures to set the adapter magnification. Display the [Main] screen. a , and select from the items displayed. Click Items to be displayed are described below. Magnification X0.5 (Initial setting) X1.0 Setting the size of image sensor This section describes the procedures to set the size of image sensor.
CKX-CCSW Setting the number of images This section describes the procedures to set the number of images to be measured. Display the [Main] screen. In [Number of Images], input the numerical value into the a field directly or click b to display the value in the left field, and set it.
[General] screen Click [General] c to display the [General] screen. Working Folder File Name Language DP Image Format Calibration About Setting the file format used for saving results This section describes the procedures to set the file format to be used for saving the results.
CKX-CCSW Setting the count color This section describes the procedures to set the color used for measuring the cell count. Display the [Setting] screen. Click the [Confluency Checker] tab a . Click b , and select from the list displayed.
Setting the confluency color This section describes the procedures to set the cell color and the background color used when measuring the confluency. Display the [Setting] screen. Click the [Confluency Checker] tab a . Click b in [Cell] and select from the list displayed. You can select from following colors.
CKX-CCSW Displaying the focus indicator The focus indicator is displayed when DP22/DP27 is connected. It shows the contrast level of the live image as a focusing guide. This section describes the procedure to display the focus indicator. Display the [Setting] screen.
Setting the working folder This section describes the procedures to set the working folder. Display the [Setting] screen. Click [General] tab a . Click The screen to select the folder appears. Select the working folder and click [Select Folder]. Setting the file name This section describes the procedures to set the file names of the acquired images or measurement results.
CKX-CCSW Setting the language This section describes the procedures to change the language of displays such as menus to different language. Display the [Setting] screen. Click [General] tab a . b , and select from the list displayed. Click The selected language is displayed in the field on the right.
Setting the calibration of DP22/DP27 If the difference in brightness or color between the right half and the left half of the image acquired using DP22/DP27 is noticeable, you can correct the image to the normal status by calibrating DP22/DP27. If DP22/DP27 is not connected, this item is not displayed.
CKX-CCSW 6-4 [Property] screen On the [Property] screen, you can confirm the contents specified in the [Main] screen and [Setting] screen. This section describes the items displayed on the [Property] screen. Clicking [ Default ] will return the settings to the initial values.
6-5 Saving the settings You can save the items displayed on the [Property] screen as the setting file. This section describes the procedures. Display the [Main] screen. Click a to display the screen to select the folder. Set the folder and the file name to save settings and click [Save].
CKX-CCSW Effective measurement of cell count/confluency 7-1 Zoom area This section describes the procedures to operate the zoom area. Item Description Click this icon to zoom out. The same operation can be performed by pinching in on the tablet PC.
Setting the cell scale bar This section describes the procedures to set the cell scale bar. Display the [Setting] screen. Click the [Confluency Checker] tab a . Tick [Display] b to display the cell scale bar. Tick [Small Cell Mode] c if you want to measure the small cell (approx.
CKX-CCSW Specifications Item Specifications Function Cell count measurement, confluency measurement Controllable camera DP22, DP27 Normal mode: 20 to 200 μm (30 to 60 μm is recommended) Small cell mode: 10 to 20 μm Measurable cell size * 130 to 200 μm: Measurable only for colony of the iPS/ES cell...
Troubleshooting Under certain conditions, performance of CKX-CCSW may not be given properly. If problems occur, please review the following list and take remedial action as needed. If the phenomena are not improved, contact Olympus for assistance. Phenomena or error message...
Page 61
CKX-CCSW Phenomena or error message Cause Remedy Page The image is not displayed when The folder where the acquired Set the folder where the acquired measuring while acquiring the image with image file is saved is not set as a...
Page 62
Or the system received the PC. If the same message still failure notification from the camera appears, contact Olympus. while this product is running. PC was in the sleep mode once. Shutdown PC once and restart PC.
Page 63
Manufactured by Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan Distributed by 48 Woerd Avenue Waltham, MA 02453, U.S.A. 8F Olympus Tower, 446 Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509 AX8857 01...
Need help?
Do you have a question about the CKX-CCSW and is the answer not in the manual?
Questions and answers