Fluke 1523 Technical Manual page 96

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1.800.544.2843
Calibration of Your Reference Thermometer Readout
4. Set all adjustment parameters to 0.0 for each range by sending the following
commands:
5. Take alignment data by following (or repeating) the steps in the As Found Data
section. Do not take alignment data for Channel T2.
6. Calculate the average and the standard deviation of the alignment data..
7. Calculate the adjustment values for each range and record the value using the
formula below:
www.calcert.com
15
10
20
5
25
0
30
Manual Calibration Process