Example
To transfer all bank data from Unit 6 to the host:
Input :@06BS̲AL[CR]
(Data will be transferred using the XMODEM protocol.)
Output :@06BS̲OK[CR]
----------------------------------------------------------------------------------------------
■ HT -Head Type- :Acquire head type
----------------------------------------------------------------------------------------------
HT acquires the head type.
Input Command
@<unit̲number>HT[CR]
・ Set the following numerical values for <unit̲number>.
Unit No.
Content
00 to 31
Unit number of the Unit whose head type is to be acquired.
Output Response
@<unit̲number>HT̲<type>[CR] :Executed correctly
<unit̲number>
<type>
Example
To acquire the head type of Unit 30:
Input: @30HT[CR]
Output: @30HT̲S15R[CR] (when F10-S15R is connected)
----------------------------------------------------------------------------------------------
■ LG -Level Get-
: Acquire threshold level
----------------------------------------------------------------------------------------------
LG acquires the threshold level. Four threshold levels are retained in
memory, regardless of the current measurement mode. The specified
level will be returned.
Input Command
@<unit̲number>LG̲<type>[CR]
・ Set the following numerical values for <unit̲number>.
Unit No.
Content
00 to 31
Unit number of Unit whose threshold level is to be acquired.
00 to 31:Designated unit number
S30R:F10-S30R
S15R:F10-S15R
S05R:F10-S05R
43
・ Set the following numerical values for <type>
Type
Content
0
Contrast threshold(plain measurement)
1
Concentration mean threshold (plain measurement)
2
Agreement level threshold(pattern measurement)
3
Measured item selection (pattern/plain) level threshold
Output Response
@<unit̲number>LG̲<level>[CR] :Executed correctly
@<unit̲number>LG̲ER[CR]
<unit̲number>
<level>
8
7.Threshold level
7
6
5
4
3
2
1.Threshold level
1
Example
To acquire the threshold level of the Pattern Measurement Mode of
Unit 5:
Input: @05LG̲2[CR]
Output: @05LG̲4[CR] (when threshold level is 4)
----------------------------------------------------------------------------------------------
■ LS -Level Set- : Set threshold level
----------------------------------------------------------------------------------------------
LS sets the threshold level. Any of the four threshold levels can be set
regardless of the current measurement mode.
Input Command
@<unit̲number>LS̲<type>̲<level>[CR]
:Not executed correctly
00 to 31: Designated unit number
1 to 7: Threshold level
44