B.13.10 Remote/Computer Interface Parameters
(There is one menu per interface.)
These two categories allow the user to configure each serial interface. Refer to Chapter 3
"Interfaces," for a thorough description of the serial interfaces and their command sets and
protocols. See table below for Computer Interface parameters.

Table B-9. Computer Interface Parameters

Function Name
(FOR RMT
CONFIGURATION
ONLY)
Device Address
Baud Rate
Parity
Protocol
Check Byte
CR Enable
LF Enable
Bad Check Byte
Loopback Test
Installation Test
D
.01023245 R
.R
OC
EV
Possible Entries
RMTPNL and USER
Numerical entry: 32
to 126
10
Toggle between 9600, 4800,
2400, 1200, 600, 300
Toggle between Odd, Even,
Mark, None
Toggle between ASCII, STX/ETX
Toggle between Long and Sum
Toggle between YES/NO
Toggle between YES/NO
Toggle between YES/NO
THIS DOCUMENT IS THE PROPERTY OF COMMUNICATIONS & POWER INDUSTRIES.
REPRODUCTION OR RELEASE WITHOUT EXPRESS PERMISSION IS STRICTLY PROHIBITED
Description
RMTPNL configures the Remote port for use
with the Remote panel. USER configures the
Remote port to mimic the CIF port's
command set and protocol capabilities. If
RMTPNL is selected, the rest of the settings
in this menu are not available.
10
If Protocol is set to STX/ETX, then Check
Byte must be Long (short for longitudinal). In
this case, Sum can't be chosen.
Press ENTER to start (sends "V"); will
display PASS if unit sees itself (receives
"V"), FAIL if it doesn't. Press ENTER again
or leave this item to stop the test. For the
unit to see itself, loop TX+ to RX+ and TX- to
RX-.
Press ENTER to send a steady stream of
"CHPA" using the user-selected interface
parameters. Press ENTER again or leave
this item to stop the test.
CHPA S
ERIES
B-21