Troubleshooting; Alarms And Warnings - Prior Lumen200 User Manual

Table of Contents

Advertisement

SECTION 7
7.1

Alarms and Warnings

The Lumen may display several warnings the following table explains the actions
recommended on hearing an alarm.
Message
Alarm reason
Bulb Fault 1
Software not
recognising hr count
from bulb
Over Temp
Bulb over temperature
Fault
Under Temp
Bulb area under
Fault
temperature
CHANGE
Bulb reach 2000hr
BULB
lifetime limit
L u m e n 2 0 0 / S / P r o

TROUBLESHOOTING

Quiet Alarm
No
Yes,
Hold button for 5-10s
Reset to clear alarm
No
Yes,
Hold button for 5-10s
Actions
Switch off unit.
Check bulb for
damage and check
data connector is
securely inserted
into its socket on
the bulb PCB.
(refer to Fig 7 page
15).
Check vents at rear
of unit are not
blocked or
covered.
Check unit is
o
above 10
C,
contact Prior.
Change bulb.
53