Connecting To The Microscope - Prior Lumen200 User Manual

Table of Contents

Advertisement

Fig 30 – The Data Connector
5.2.2

Connecting to the Microscope

Once the bulb is installed (see section 5.2), the Lumen200Pro is ready to be connected to the
microscope.
1. Place the Lumen200Pro in a safe location and ensure none of the fan outputs are obstructed.
2. Unpack the liquid light guide from the foil packaging and remove both plastic caps from the
light guide. Note: It is important to remove the caps before connecting to the light guide to
avoid damage on power up.
3. Unscrew the connector on the front of the Lumen200Pro, and insert the Light guide.
a. Ensure the light guide is fully inserted into the Lumen200Pro, the light guide should
be inserted by 58mm.
4. Tighten the connector unit resistance is felt, the light guide is now firmly held in the
Lumen200 Pro connector.
5.
Locate the collimating lens supplied to attach the light guide onto the microscope, loosen
the screw on the back of the collimating lens. Firmly push the light guide into the hole
ensuring it has reached it end stop and tighten the screw.
a.
The Lumen200Pro is supplied with a collimating lens specifically designed to fit
your microscope, the supplied collimating lens will only fit the specified make of
microscope.
38
Data Connector
P r i o r S c i e n t i f i c

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lumen200sLumen200pro

Table of Contents