4.7.2 General Commands for the Lumen 200S when connected via OptisScan / ProScan
Controllers
Command
Argument s
8
s,c[,t]
8
s
Shutter
s
L u m e n 2 0 0 / S / P r o
Response
Description
(All end with <cr>)
R
Opens or closes the shutter s (value '1'
'2' or '3'), if c is 0 the shutter is opened,
If shutter s is not
fitted E,20 will
1 it is closed. The optional argument t is
be returned
used to open/close the shutter for a
time t milliseconds.
c
Returns status c of shutter s
If shutter s is not
fitted E,20 will
be returned
Text string
Prints information about shutter's' (s is a
value between 1 and 3). The information
end is always a line saying END. This
allows for the addition of extra fields of
information without effecting
application software. Users should
always read lines until the END is seen in
order to maintain compatibility.
Example
SHUTTER_1 = NORMAL
DEFAULT_STATE=CLOSED
END
SHUTTER_2=NONE
END