4 . 7 RS232 Command Set for Lumen200S

A description of how to connect to the ProScan or OptiScan systems is supplied in the
OptiScan II and ProScan II manual. The following is the RS232 commands applicable to
the Lumen 200Pro.
4.7.1 General Commands for the Lumen 200S when connected to the PC
Command
Arguments
?
None
VER
None
VERSION
None
L u m e n 2 0 0 / S / P r o
Response
Description
(All end with <cr>)
Text string
Reports information about the peripherals currently
connected to the controller. The information end is always
a line saying END This allows for the addition of extra fields
of information without effecting application software.
Users should always read lines in until the END is seen.
A typical response is from Lumen 200S shown below
PROSCAN INFORMATION
DSP_1 IS NOT FITTED
DSP_2 IS 3 AXIS STEPPER VERSION 0.0
DRIVE CHIPS 111000
JOYSTICK NOT FITTED
STAGE = NONE
FOCUS = NONE
FOURTH = NONE
FILTER_1 = NONE
FILTER_2 = NONE
SHUTTERS = 001
LED = 0000
AUTOFOCUS = NONE
VIDEO = NONE
HARDWARE REV A
END
Text string
Returns the version of the code fitted to the unit.
Eg. LumenShutter V2.5c
Text string
Returns the Issue number.
Eg. 025