APPENDIX E – COMPATIBILITY WITH THE OPTISCAN III ..........45 APPENDIX F – INFORMATION FOR CUSTOMERS IN CHINA ........... 46 Many thanks for purchasing an OptiScan III – we hope and expect that it will prove to be useful, reliable and an asset to your microscopy work.
This manual covers all varieties of the OptiScan III controller produced by Prior Scientific. It also covers the control, via the OptiScan III, of any Prior product which is compatible with the OptiScan III, including the control joystick, focus motors and the stepper motor stage.
3.2 Fitting the OptiScan III Stage The Prior OptiScan III stage is supplied with a base plate insert to suit the microscope specified. Place the stage onto the microscope stage mount and attach using the fixing crews or clamping screw supplied.
3.3 Cable Connections The cable connections to the OptiScan III controller are located on the rear panel of the control box, as shown in the illustration below. Before making any of these connections, ensure that the OptiScan III controller is switched off. Each connection is labelled. Do not connect your computer's serial port cable to the `Z' axis connector on the controller, this may damage your computer and the OptiScan III controller.
If nothing happens then try changing the coms port (using the box in the left hand corner of Prior Terminal’ and again type ‘?’. If there are a large number of coms ports, or for any other reason you don’t wish to use this method, open the control panel from the start menu, open “system”...
4.2 Using Joysticks (CS152DP and CS200XY and CS200Z) CS152DP Joystick The joystick unit provided as part of the OptiScan III system is a 3 axis joystick (CS152DP) used to control the motorized stage and the focus motor. CS152DP joystick – 3axis, controls stages and focus The joysticks feature an X, Y joystick, two sliding tensioners and two ‘Hot Keys’.
OPTISCAN III MANUAL CS200XY joystick and CS200Z focus The joystick unit provided as part of the Optiscan III system is the CS200XY joystick and the CS200Z focus are used to control the Optiscan stage and the focus motor. Connect to...
1mm in X. The STAGE, and FOCUS command responses are terminated with the word END. This will enable extra information about the OptiScan III to be added in the future and still be readable by the Application Software.
5.4 Differences between the OptiScan III and OptiScan II Several hardware and software changes have been made between the two generations of the OptiScan controller. These software changes now allow the OptiScan III and ProScan III to use the same command structure.
Y = D01 X = D00 OptiScan III treats stage (X and Y axis) as a single resource and as such a move in Y or X will only return 3. (X and Y are kept separate for compatibility with ProScan.
Page 16
Allowable values for baud are 9600 (b=96) 19200 (b=19), 38400 (b=38) or 115200 (b=15). WARNING: If Baud rate of the OptiScan III is changed it is important for Application software to check communication with OptiScan III by scanning Baud rate on initialisation.
For Installation see section 4. Check that the controller is switched on. Check that the green “Running” light is on. If it is not, return unit to your local Prior Scientific dealer for servicing. There is no external fuse that can be replaced.
Page 27
System will not respond to remote RS232 commands Suggestions: Set Up Prior Terminal (see Appendix C). Check that the RS232 cable is plugged into the RS232- 1 port in the back panel of the OptiScan II Controller. Press “Enter” on your computer keyboard several times.
Return Material Authorisation (RMA) number must be obtained from the appropriate Prior Scientific office before returning any equipment. For North and South America contact Prior Scientific Inc. and for the rest of the world call Prior Scientific Instruments Limited on the telephone numbers shown below.
OPTISCAN III MANUAL APPENDIX D: Installing and Using the software Prior Scientific has developed its own software allowing the control of Prior products via USB or RS232 commands. Although many users will wish to incorporate our equipment into their own programs, using Prior software means that our products will perform at their peak, and that diagnosing the causes of problems may be easier.
Page 41
Connect your product to the computer via a USB cable. Prior Terminal Click on the Windows Start menu and go to ‘All Programs’. Go to ‘Prior Scientific’ and select ‘Visual Basic’. From this, go to ‘Prior Terminal’ and open that program. Other programs are also available.
Page 43
9600; however some software can change this. A successful connection is indicated in the Status Bar of the Prior Terminal Program. Once Prior Terminal is open, to list what hardware is connected type ‘?’. From the product manual you should see the correct response this should generate.
Page 44
Number is the one under “USB serial port”. Ensure that this is the port selected by Prior Terminal by clicking the ‘Com’ button. If this fails to resolve the problem then it is probably the case that the driver is not installed properly.
OPTISCAN III MANUAL APPENDIX E – COMPATIBILITY WITH THE OPTISCAN III The following products are compatible with/require the OptiScan III system and are covered by the declaration of conformity. An * denotes that an additional manual is available and should be read in addition to this manual in order to successfully use the system.
Need help?
Do you have a question about the OptiScan III and is the answer not in the manual?
Questions and answers