Starting Up The Lumen200Pro - Prior Lumen200 User Manual

Table of Contents

Advertisement

7. Using the Prior Filter block tool unscrew the filter ring, making sure to locate the two lugs
in the filter block tool into the two holes of the filter ring.
8. Remove the old filter; replace the new filter taking care not to place finger prints on the
filter.
9. Using the Prior Filter block tool screw the filter ring to hold the filter. Be careful; not to
over tighten the filter ring; some filters will expand when heated by the light.
10. Repeat for all required filters. Remember to record the position of each filter for use with
your software.
11. Replace top of unit, ensuring the internal connector is aligned and replace screws.
12. Reconnect controller, check that the filter wheel is moving.
13. Reconnect Lumen200Pro mains power.

Starting Up the Lumen200Pro

Warning:
• Do not power up the lumen without the light guide attached to both the
Lumen200Pro and microscope.
• Only power up the Lumen200Pro when it is installed on a level surface.
1 Ensure the light guide is attached to both the Lumen200Pro and microscope.
2 Connect the power cable to the Lumen200Pro.
3 Switch the Lumen200Pro power switch on.
4 Allow approximately 5 minutes for light to reach operational temperature.
5 Do not power down the unit within 10 mins of power up. This may reduce the effective
lifetime of the bulb.
L u m e n 2 0 0 / S / P r o
43

Advertisement

Table of Contents
loading

Related Products for Prior Lumen200

This manual is also suitable for:

Lumen200sLumen200pro

Table of Contents