Fig 7 - The Data Connector
3.2.2

Connecting to the Microscope

Once the bulb is installed (see section 3.2.1), the Lumen200 is ready to be connected to the
microscope.
1. Place the Lumen200 in a safe location and ensure none of the fan outputs are obstructed.
2. Unpack the liquid light guide from the foil packaging and remove both plastic caps from the
light guide. Note: It is important to remove the caps before connecting to the light guide to
avoid damage on power up.
3. Unscrew the connector on the front of the Lumen200, and insert the Light guide.
a. Ensure the light guide is fully inserted into the Lumen200, the light guide should be
inserted by 58mm.
4. Tighten the connector unit resistance is felt, the light guide is now firmly held in the
Lumen200 connector.
L u m e n 2 0 0 / S / P r o
Data Connector
15