8 Function group "output" (06), - "profibus param." (06), PROFIBUS-PA only
8.9

Function "fixed cur. value" (064), HART only

Set the fixed current value with this function. This data is necessary if you have selected
the "fixed current" option in the "curr. output mode" (063) function.
User input:
3,8...20,5 mA
Caution!
This user input is available for HART devices only!
8.10 Function "out status" (064), PROFIBUS-PA only
Displays the current output status (for value, see operating instructions of relevant
instrument).
Caution!
This function is available for PROFIBUS-PA devices only!
51