Endress+Hauser levelflex M FMP 40 Manual page 17

Hide thumbs Also See for levelflex M FMP 40:
Table of Contents

Advertisement

Levelflex M
Endress+Hauser
FMP41C, FMP45:
ε
Media
DC (
r)
group
0
unknown
1
1,4 ... 1,6 – Liquefied gases, e.g. N
2
1,6 ... 1,9 – Liquefied gas, e.g. propane
– Solvents
– Frigen / freon
– Palm oil
3
1,9 ... 2,5 – Mineral oils, fuels
4
2,5 ... 4
– Benzene, styrene, toluene
– Furan
– Naphthalene
5
4 ... 7
– Chlorobenzene, chloroform
– Cellulose spray
– Isocyanate, aniline
6
> 7
– Aqueous solutions
– Alcohols
– Ammoniac
The lower group applies to very loose or loosened bulk solids.
Reduction of the max. possible measuring range by means of:
• extremely loose surfaces of bulk solids, e.g. bulk solids with low piled density when
filled pneumatically.
• Build-up, primarily of moist products.
3 Function group "basic setup" (00)
Typical liquids
Typical measuring
range FMP41C
, CO
4 m when installing
2
2
in metallic pipes
Typical measuring
range FMP45
4 m coax probe and
when installing in
metallic pipes
9 m
25 m
12 m
30 m
16 m
35 m
25 m
35 m
35 m
35 m