Chapter 8 - Test and Inspection Matrix
transducer touches the flat surface only with the button and that the
transducer is parallel to the flat surface.
8. The Toco display should read between 20 and 40 units above the
Baseline.
Toco display = 5, 10, 15, or 20
depending on Baseline
setting
Figure 12
Testing a Toco Transducer
9. If the test fails, repeat with another transducer. If it still fails, refer to
Chapter 5, "Troubleshooting".
Testing Transducers
Toco display = 20 - 40 Toco
units above the Baseline
91