Troubleshooting the Pump Leak Rate Test

Secondary Leak > 3 µL/min
• Leak between the OBV and automatic purge valve
• Outlet valve not properly assembled
Dynamic Leak > 30 µL/min or Dynamic Leak Rate Test fails
• Air in the primary pump chamber
• Abort due to over pressure
• Leak in Inlet Valve
1290 Infinity II High Speed Pump User Manual
Check for visible leaks on fittings and connectors
Connector not fixed / tight enough
Connector damaged
Leaky filter frit assembly
Remove the seal wash tubes from the support ring and check for leak
into the seal wash path
Main seal leaking/damaged
Piston damaged
Remove waste lines from the automatic purge valve
Damage to rotor seal and/or stator head
Re-tighten the outlet valve
Check the position of the gold seal
Check for air in the solvent inlet lines and the Tuning signal
Purge the lines, Prime and Condition the pump head
Check solvent and solvent settings
Purge and condition the pump head with water
Check for moving air bubbles in tubing directly to the Inlet Valve
Purge the lines with water to remove dirt
Knock at the valve, clean or replace it