Troubleshooting and Diagnostics
Pump Leak Rate Test
Figure 12

Troubleshooting the Pump Leak Rate Test

Secondary Leak > 3 µL/min
• Leak between the OBV and automatic purge valve
84
Dynamic (primary) Leak Test
Check for visible leaks on fittings and connectors
Connector not fixed / tight enough
Connector damaged
Leaky filter frit assembly
Remove the seal wash tubes from the support ring and check for leak
into the seal wash path
Main seal leaking/damaged
Piston damaged