Resistance And Continuity Function Tests; Low Battery Indicator Test; Parts And Accessories - Fluke T5-600 Service Information

Hide thumbs Also See for T5-600:
Table of Contents

Advertisement

T5-600/T5-1000
Electrical Tester

Resistance and Continuity Function Tests

1. Set the calibrator to the resistance given in Step 1 of Table 5.
2. Apply the resistance to the tester. Verify that the tester reads within the display limits shown.
3. Apply the resistances and verify the tester's responses for the remaining steps in Table 5.
Step
1
2
3
4

Low Battery Indicator Test

The following procedure verifies correct operation of the low battery indicator.
1. Remove the tester's batteries.
2. Set the dc power supply to 3.0 V. Apply this voltage to the tester's battery terminals.
3. Set the calibrator to 1 kΩ. Apply this resistance to the tester's probes.
4. Slowly decrease the dc voltage just until the tester's low battery indicator (M) turns on.
5. Verify that the tester reads 1000 ±12 counts (988 to 1012).
6. Disconnect the calibrator and the dc power supply. Reinstall the tester's batteries.

Parts and Accessories

Tables 6 shows the replacement parts and accessories available from Fluke for the T5-600 and T5-1000
Electrical Testers.
Test lead assembly
Replace only with Fluke double-insulated leads.
Battery door
Battery door screw
AA battery, 1.5 V, carbon-zinc (2 required)
AA battery, 1.5 V, alkaline (2 required)
T5-600/T5-1000 Electrical Tester Instruction Sheet (English)
8
1.800.561.8187
Table 5. Resistance and Continuity Tests
Calibrator Output
1 kΩ
26 Ω
300 Ω
0 Ω
Table 6. Replacement Parts and Accessories
Description
www.
T5-600/T5-1000 Display Limits and Beeper Responses
988 to 1012
Beeper ON
Beeper OFF
-2 to 2, Beeper ON
information@itm.com
.com
Fluke Part Number
648029
1626588
1618578
650181
376756
1629509

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T5-10007956000079559999095969946759095969946803

Table of Contents